A-96.250.211 / 171121
AMI Sodium P
Program List and Explanations
106
5.3.2 and 5.3.3
Relay 1 and 2:
The contacts can be set as normally open or
normally closed with a jumper. See
The function of relay contacts 1 or 2 are defined by the user.
Note:
The navigation in the menu <Relay 1> and <Relay 2> is
equal. For reason of simplicity only the menu numbers of Relay 1
are used in the following.
1
First select the functions as:
- Limit upper/lower,
- Control upwards/downwards,
- Timer
- Fieldbus
- Channel Selection (relay 2 only)
2
Then enter the necessary data depending on the selected
function.
When the relays are used as upper or lower limit switches, program
the following:
5.3.2.20
Parameter:
select a process value
5.3.2.300
Setpoint
: If the measured value rises above respectively falls below
the set-point, the relay is activated.
5.3.2.400
Hysteresis:
within the hysteresis range, the relay does not switch.
This prevents damage of relay contacts when the measured value
fluctuates around the alarm value.
5.3.2.1
Function = Limit upper/lower:
Parameter
Range
Sodium 1
0–20 ppm
Sodium 2
0–20 ppm
pH
0–14 pH
Temperature
-30 to +120 °C
Sample flow
0–1000 B/s
Parameter
Range
Sodium 1
0–20 ppm
Sodium 2
0–20 ppm
pH
0–14 pH
Temperature
0 to +100 °C
Sample flow
0–1000 B/s