A-96.250.211 / 171121
AMI Sodium P
Program List and Explanations
105
5.3.1.4.1
Flow Alarm:
Program if the alarm relay should be activated if there is
a flow alarm. Choose between yes or no. The flow alarm will always
be indicated in the display, pending error list, saved in the message
list and the logger.
Available values: Yes or No
Note:
Sufficient flow is essential for a correct measurement.
We recommend to program yes.
5.3.1.4.2
Alarm High:
If the measuring values rises above the programmed
value E009 will be issued.
Range: 5–1000 B/s
5.3.1.4.35
Alarm Low:
If the measuring values falls below the programmed val
-
ue E010 will be issued.
Range: 5–1000 B/s
5.3.1.5
Sample Temperature:
Define the measuring value, which should is
-
sue an alarm high respectively low.
5.3.1.5.1
Alarm High:
If the sample temperature rises above the programmed
value E007 is issued.
Range: 30–70 °C
5.3.1.5.x
Alarm Low:
If the sample temperature falls below the programmed
value E008 is issued.
Range: 0–20 °C
5.3.1.6
Case Temp. high:
Set the alarm high value for temperature of elec
-
tronics housing. If the value rises above the programmed value E013
is issued.
Range: 30–75 °C
5.3.1.7
Case Temp. low:
Set the alarm low value for temperature of electron
-
ics housing. If the value falls below the programmed value E014 is
issued.
Range: -10–20 °C