A-96.250.211 / 171121
AMI Sodium P
Program List and Explanations
96
3.2 Regeneration
Note:
The Regeneration menu is only visible if the automatic
regeneration option is installed.
3.2.1
Manual operation:
Allows to manually open and close the solenoid
valve to test it or to fill the tube with regeneration solution.
3.2.2
Parameters:
3.2.2.1
Start time:
The programmed start time is valid for each of the pro
-
grammed days.
3.2.2.2
Calendar:
3.2.2.3
Duration:
Duration of the automatic sensor regeneration.
Range: 10–60 sec
3.2.2.4
Delay:
During regeneration plus delay time, parameters
Range: 0–6‘000 Sec
3.2.2.5
Signal Outputs:
Select the operation mode of the signal outputs
during regeneration:
3.2.2.6
Output/Control:
Relay or signal output:
3.2.2.2.1
Monday: Possible settings: on or off
to
3.2.2.2.7
Sunday: Possible settings: on or off
Cont.:
Signal outputs continue to issue the measured value.
Hold:
Signal outputs hold the last valid measured value. Errors,
except fatal errors, are not issued.
Off:
Signal outputs are switched off (set to 0 or 4 mA).
Errors, except fatal errors, are not issued.
Cont.:
Controller continues normally.
Hold:
Controller continues based on the last valid value.
Off:
Controller is switched off.