A-96.250.211 / 171121
AMI Sodium P
Maintenance
71
6.7.1
Cleaning the Flow cell
Disassemble
the flow cell
1
Shut off the instrument.
2
Close the sample flow main tap.
3
Drain the flow cell [D] completely.
4
Remove all sensors [A].
5
Remove all tube connections.
6
Unscrew the 4 screws [B] of the flow cell cover [C] and remove it.
7
Clean the flow cell with a soft brush.
Assemble the
flow cell
1
Screw the cover onto the flow cell.
2
.
3
.
4
Open sample flow main tap.
5
Switch on the instrument.