AMI Sodium P
Program List and Explanations
97
3.3 Simulation
To simulate a value or a relay state, select the
alarm relay,
relay 1 or 2
signal output 1 or 2
valve 1
with the [
] or [
] key.
Press the [Enter] key.
Change the value or state of the selected item with the [
] or
[
] key.
Press the [Enter] key.
The value is simulated by the relay/signal output.
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated val
-
ues will be reset.
3.4 Set Time
Adjust date and time.
4 Operation
4.1 Sensors
4.1.1
Filter Time Constant:
Used to damp noisy signals. The higher the fil
-
ter time constant, the slower the system reacts to changes of the
measured value.
Range: 5–300 sec
4.1.2
Hold after Cal:
Delay permitting the instrument to stabilize again after
calibration. During calibration- plus hold-time, the signal outputs are
frozen (held on last valid value), alarm values, limits are not active.
Range: 0–6‘000 sec
3.3.1
3.3.2
3.3.3
3.3.4
3.3.5
3.3.6
3.3.7
Alarm Relay:
Relay 1:
Relay 2
Signal Output 1:
Signal Output 2
Signal Output 3:
Solenoid valve
Active or inactive
Active or inactive
Active or inactive
Actual current in mA
Actual current in mA
Actual current in mA (option)
Not available.