A-96.250.211 / 171121
AMI Sodium P
Program List and Explanations
98
4.1.31
Interval:
Only visible if <Channel switch> in menu 5.1.4 is set to
<Auto>. The measuring interval can be set in steps of 15 min.
Range: 15 min to 120 min
4.1.32
Channel selection:
Only visible if <Channel switch> in menu 5.1.4 is
set to <User defined>. The following selections are available:
4.2 Relay Contacts
See
5.3 Relay Contacts, p. 103
.
4.3 Logger
The instrument is equipped with an internal logger. The logger data
can be copied to a PC with an USB stick if option USB interface is in
-
stalled.
The logger can save approx. 1500 data records. Records consists of:
Date, time, alarms, measured value, pH, temperature, flow.
4.3.1
Log Interval:
Select a convenient log interval. Consult the table be
-
low to estimate the max logging time. When the login buffer is full,
the oldest data record is erased to make room for the newest one
(circular buffer).
4.3.2
Clear Logger:
If confirmed with
yes
, the complete logger data is de
-
leted. A new data series is started.
Channel 1:
Only channel 1 is measured.
Channel 2:
Only channel 2 is measured.
Dig. Input:
The channel can be selected via input. The function
Input in menu 5.3.4 is set to <Active = no>.
Interval
1 s
5 s
1 min
5 min
10 min
30 min
1 h
Time
25 min
2 h
25 h
5 d
10 d
31 d
62 d