A-96.250.211 / 171121
AMI Sodium P
Default Values
115
Scaling: pH: Range high: ......................................................14.00 pH
Scaling: Sample Flow: Range low: .............................................0 B/s
Scaling: Sample Flow: Range high:...................................... 1000 B/s
Alarm Relay
Alarm Sodium:
Alarm high:......................................................................... 20.00 ppm
Alarm low:............................................................................. 0.00 ppb
Hysteresis: ............................................................................ 10.0 ppb
Delay:............................................................................................. 5 s
pH: Alarm high: .....................................................................14.00 pH
pH: Alarm low:.........................................................................0.00 pH
pH: Hysteresis: .......................................................................0.10 pH
pH: Delay:...................................................................................... 5 s
Flow Alarm:................................................................................... Yes
Sample flow.: Alarm High:.................................................... 1000 B/s
Sample flow.: Alarm Low:........................................................... 5 B/s
Sample Temp.: Alarm High: ....................................................... 55 °C
Sample Temp.: Alarm Low: .......................................................... 5 °C
Case temp. high:........................................................................ 65 °C
Case temp. low:........................................................................... 0 °C
Relay 1 and 2
Function:........................................................................... Limit upper
Parameter: Relay1 and 2.....................................................Soduim 1
Setpoint: Relay1 and 2 ........................................................ 1.00 ppm
Hysteresis: ............................................................................... 10 ppb
Delay:........................................................................................... 30 s
If Function = Control upw. or dnw:
Parameter:
................................................................Sodium 1 and 2
Settings: Actuator: ............................................................. Frequency
Settings: Pulse Frequency:.................................................... 120/min
Settings: Control Parameters: Setpoint:........................... 1.00 ppm
Settings: Control Parameters: P-band:................................. 10 ppb
Parameter:
..................................................................................... pH
Settings: Actuator: ............................................................. Frequency
Settings: Pulse Frequency:.................................................... 120/min
Settings: Control Parameters: Setpoint:..............................7.00 pH
Settings: Control Parameters: P-band:................................0.10 pH
Parameter:
.....................................................................Temperature
Settings: Actuator: ............................................................. Frequency
Settings: Pulse Frequency:.................................................... 120/min
Settings: Control Parameters: Setpoint:................................. 30 °C
Settings: Control Parameters: P-band:..................................... 1 °C