A-96.250.211 / 171121
AMI Sodium P
Product Description
14
Second
Sample Stream
If required the AMI Sodium P can be equipped with the optional sec
-
ond sample stream module. To install the 2
nd
sample stream option,
an AMI Sodium P with 400 mm or 375 mm panel width is required.
Sample
Sequencer
If measurement of more than two sample streams is required, the
AMI Sodium P can be connected to a Sample Sequencer, which al
-
lows to measure up to six sample streams.
Automatic
Regeneration
Optionally, the AMI Sodium P can be equipped with a module for
automatic sensor regeneration. To install the automatic regeneration
module, an AMI Sodium P with 400 mm panel width is required.
Grab Sample
The standard bottle holder can also be used for a grab sample mea
-
surement. For details, see
Calibration
A standard bottle is screwed in the standard bottle holder [Q] and
turned upwards into the vertical position, thus switching the sample
flow from the constant head to the standard bottle. Constant pres
-
sure within the standard bottle is maintained by the pressure-equaliz
-
ing tube inside the bottle.
One liter of standard is consumed in approx. 10 min. The sodium
electrode must reach constant readings within this time to obtain an
exact calibration.
For details, see
Calibration, p. 42.
Consumables
The filling of the 100 ml KCl bottle lasts for one month of operation.