AMI Sodium P

Instrument Setup

50

4.

Instrument Setup

4.1. Installing the Reagent Bottle

CAUTION
Formation of reagent vapor

To prevent formation of reagent vapors:

close the reagent bottle firmly

check the EPDM seal regularly

install air tubes and filter properly

Note: 

Operate the instrument only with diisopropylamine.

A

B

C

D

E

F

G

H

Screw cover G 45

Tube 1 from air filter

Tube 2 to reaction tube

Tube holder

EPDM seal

Air inlet tube

Reagent bottle

Bottle holder

A

B
C

E

F

D

H

G

Содержание AMI Sodium P

Страница 1: ...Operator s Manual Firmware V6 22 and higher AMI Sodium P A 96 250 211 171121...

Страница 2: ...n representative or the manufacturer Swan Analytische Instrumente AG Studbachstrasse 13 8340 Hinwil Switzerland Internet www swan ch E mail support swan ch Document Status Title AMI Sodium P Operator...

Страница 3: ...2 Installing the Reference Electrode 27 3 4 3 Installing the pH Electrode 31 3 4 4 Installing the Temperature Sensor 32 3 4 5 Installing the reagent bottle 32 3 5 Installing the 2nd Sample Stream Opt...

Страница 4: ...nce of Reference Electrode 64 6 5 Maintenance of pH Sensor 65 6 6 Maintenance of Solenoid Valve 66 6 6 1 2nd Sample Stream Option 66 6 6 2 Automatic Regeneration Option 68 6 7 Maintenance of Flow Cell...

Страница 5: ...90 8 3 Maintenance Main Menu 3 91 8 4 Operation Main Menu 4 92 8 5 Installation Main Menu 5 92 9 Program List and Explanations 94 1 Messages 94 2 Diagnostics 94 3 Maintenance 95 4 Operation 97 5 Inst...

Страница 6: ...ructions are given throughout this manual at the re spective locations where observation is most important Strictly fol low all safety instructions in this publication Target audience Operator Qualifi...

Страница 7: ...tion instructions carefully WARNING Severe injuries or damage to the equipment can occur if such warnings are ignored Follow the prevention instructions carefully CAUTION Damage to the equipment minor...

Страница 8: ...171121 AMI Sodium P Safety Instructions 8 Warning Signs The warning signs in this manual have the following meaning Electrical shock hazard Corrosive Harmful to health Flammable Warning general Atten...

Страница 9: ...li ty for any claim resulting from unauthorized modification or alter ation WARNING Electrical Shock Hazard If proper operation is no longer possible the instrument must be disconnected from all power...

Страница 10: ...tion you must read and understand the instructions in this manual as well as the Material Safety Data Sheets MSDS Etching kit for sodium electrodes powder liquid Regeneration solution for sodium elect...

Страница 11: ...ure With the current temperature the measuring value is expressed for the standard temperature of 25 C by using programmed tempera ture compensation curves Sodium measurements below 1 ppb require a sp...

Страница 12: ...overflow tube M into the waste This adjustment ensures a sufficient sample flow through the flow cell K First the sample flows into the measuring chamber K1 where the sodium electrode G is installed...

Страница 13: ...ctrode Sodium electrode Temperature sensor Bubble counter Sample with air bubbles Flow cell L M N O P Q R S T U Sample outlet Overflow tube Air inlet filter Air tube to DIPA bottle Automatic regenerat...

Страница 14: ...the automatic regeneration module an AMI Sodium P with 400 mm panel width is required Grab Sample The standard bottle holder can also be used for a grab sample mea surement For details see Grab Sampl...

Страница 15: ...num with a protection degree of IP 66 NEMA 4X 10 to 50 C 30 to 85 C 10 90 rel non condensing backlit LCD 75 x 45 mm Sample requirements pH value Ammonium concentration Flow rate Temperature Inlet pres...

Страница 16: ...ription 16 Dimensions Panel Dimensions Screws Weight stainless steel 280 x 850 x 200 mm 8 mm diameter 9 0 kg 19 85 lbs without sample water Exit Enter AMI Sodium P 13 mm 0 51 254 mm 10 00 280 mm 11 02...

Страница 17: ...imensions with 2nd sample stream Panel Dimensions Screws Weight stainless steel 400 x 850 x 200 mm 8 mm diameter 12 0 kg 26 5 lbs without sample water 13 mm 0 51 4 x dia 10 mm 0 39 850 mm 33 46 824 mm...

Страница 18: ...compact panel version Panel Dimensions Screws Weight stainless steel 375 x 700 x 200 mm 8 mm diameter 9 0 kg 19 85 lbs without sample water Exit Enter AMI Sodium P 10 mm 0 39 17 5 mm 0 69 30 mm 1 18 4...

Страница 19: ...nce electrolyte bottle 2nd sample stream option Flow regulating valve Sample inlet pH electrode Sodium electrode Bubble detector J K L M N O P Q Air filter Constant head Standard bottle Grab sample St...

Страница 20: ...lectrode Rinse it well and check for air bubbles inside the electrode Install the sodium electrode Connect cable S to the sodium electrode Reference electrode Installing the Reference Electrode p 27 I...

Страница 21: ...and fill the tube with regeneration solution by calling the Maintenance Regenerations Manual Operation menu item Instrument setup Program all parameters for external devices interface recorders etc P...

Страница 22: ...t must only be installed by trained personnel Mount the instrument in vertical position For ease of operation mount it so that the display is at eye level For the installation a kit containing the fol...

Страница 23: ...ting Use plastic tube 4 x 6 mm to connect the sample line Mounting of SERTO fitting 3 3 2 Sample Outlet WARNING Risk of water pollution The drain of the flow cell outlet contains diisopropylamine DIPA...

Страница 24: ...o the flow cell after mounting of the instrument panel The electrodes are pro tected with protective caps at their tips as well as on the electrical connectors Remove the connector caps from the conne...

Страница 25: ...re that the sensing glass bulb stays clean no air bubbles are trapped in the glass bulb of the electrode the electrical connectors stay clean and dry The electrode is delivered with protective caps on...

Страница 26: ...he O ring H and slip it carefully over the electrode shaft C 6 Make sure no air bubbles are trapped in the sensing glass bulb E Shake the electrode like a clinical thermometer with its tip pointing do...

Страница 27: ...ain correct operation make sure that the ground joint diaphragm stays clean and a KCl flow of about 1ml day is maintained no air bubbles are trapped in the electrode and in the tube to the KCl reservo...

Страница 28: ...2 Hold the reference electrode with the ground joint diaphragm pointing downwards 3 Slightly lift the ring shaped sleeve of the ground joint diaphragm and allow a little electrolyte to flow out into...

Страница 29: ...ect the KCl supply tube to the dosing tip C of the KCl bottle the compact panel version requires an additional extension tube see Extension tube on compact panel version p 30 A B C D E F KCl bottle Tr...

Страница 30: ...it carefully over the electrode shaft I 8 Insert the electrode through the hole H into the reference cham ber and push it down until the ground joint diaphragm is about 0 5 cm above the bottom 9 Tight...

Страница 31: ...y over the electrode shaft 5 Insert the electrode through the hole H in into the flow cell 6 Tighten the union screw E hand tight 7 Remove the connector cap A 8 Screw the connector B of the cable mark...

Страница 32: ...ure sensor proceed as follows 1 Remove the temperature sensor A from the panel 2 Insert the temperature sensor through the hole E into the flow cell 3 Push it into the hole as far as it will go 3 4 5...

Страница 33: ...alve from the constant head support A with the blind plug B 4 Remove the existing sample inlet tube J from the constant head support inlet 5 Screw the 2nd sample stream option C to the panel using the...

Страница 34: ...Electrical connection WARNING Electrical shock hazard Before opening the AMI transmitter switch power off Use one of the PG7 cable glands to feed the cable of the solenoid valve into the AMI transmitt...

Страница 35: ...connected to the AMI Sodium P which allows to measure up to six sample streams The electrical connection is de scribed in the manual of the AMI Sample Sequencer Note If the AMI Sodium P is already equ...

Страница 36: ...ntilation tube D to the bottle holder of the auto matic regeneration module 6 Screw the automatic regeneration module to the panel using the supplied screws if necessary unscrew the air filter C and s...

Страница 37: ...A 96 250 211 171121 AMI Sodium P Installation 37 Exit Enter AMI Sodium P G E D C F...

Страница 38: ...the cable of the solenoid valve into the AMI transmitter housing 1 Remove the plug A from the cable gland B 2 Open the transmitter housing 3 Feed the cable of the solenoid valve through the cable glan...

Страница 39: ...erate the instrument from an power outlet which has a ground connection Make sure the power specification of the instrument corre sponds to the power on site Cable thicknesses In order to comply with...

Страница 40: ...ure that the devices connected to the following contacts are disconnected from the power before resuming installation relay 1 relay 2 alarm relay WARNING To prevent from electrical shock do not connec...

Страница 41: ...tion 41 3 8 1 Connection Diagram CAUTION Use only the terminals shown in this diagram and only for the mentioned purpose Use of any other terminals will cause short circuits with possible correspondin...

Страница 42: ...o be connected to the grounding terminal Installation requirements The installation must meet the following requirements Mains cable to comply with standards IEC 60227 or IEC 60245 flammable rating FV...

Страница 43: ...e Max load 1 A 250 VAC Alarm output for system errors Error codes see Troubleshooting p 84 Note With certain alarms and certain settings of the AMI transmitter the alarm relay does not switch The erro...

Страница 44: ...us of the relays described below For programming see 5 3 2 and 5 3 3 p 106 menu Installation Relay config Terminals Jumper pos Description Relay configuration Normally Open 6 7 Relay 1 8 9 Relay 2 Ina...

Страница 45: ...n paral lel to the load A snubber circuit is not necessary if an AMI relaybox is used Resistive load Resistive loads max 1 A and control signals for PLC impulse pumps and so on can be connected withou...

Страница 46: ...ator Signal output 1 Terminals 14 and 13 Signal output 2 Terminals 15 and 13 For programming see 5 2 Signal Outputs p 99 3 11 Interface Options The slot for interfaces can be used to expand the functi...

Страница 47: ...esponding installation instruction Note Max burden 510 Third signal output 0 4 20 mA PCB 3 11 2 Profibus Modbus Interface Terminal 37 PB Terminal 38 PA To connect several instruments by means of a net...

Страница 48: ...nsult the HART manual HART Interface PCB 3 11 4 USB Interface The USB Interface is used to store Logger data and for Firmware up load For detailed information see the corresponding installation in str...

Страница 49: ...tallation 49 3 12 Connection of Sensors Front end board A B C D Temperature sensor cable marked with T Terminal 19 20 pH electrode cable marked with pH Reference electrode cable marked with R Sodium e...

Страница 50: ...ent formation of reagent vapors close the reagent bottle firmly check the EPDM seal regularly install air tubes and filter properly Note Operate the instrument only with diisopropylamine A B C D E F G...

Страница 51: ...he screw cover A onto the DIPA bottle and tighten it firmly 4 Screw the tube fitting of tube 1 B into the tube holder D so that it is connected to the air inlet tube 5 Screw the tube fitting of tube 2...

Страница 52: ...ns recirculate it into the water system 1 Swing down the bottle holder D as far as it will go 2 Open the flow regulating valve A 3 Adjust the sample flow so that a small part of the sample over flows...

Страница 53: ...annel switching mode See 5 1 4 p 99 for details Automatic regeneration option If the automatic regeneration option is installed configure the time and days of the week on which an automatic regenerati...

Страница 54: ...nges to move back to the previous menu level B to move DOWN in a menu list and to decrease digits C to move UP in a menu list and to increase digits D to open a selected sub menu to accept an entry Ex...

Страница 55: ...s 2 channel operation the active channel is flushed measurement of active channel 1 channel operation In 1 channel operation no symbols are displayed F Sample temperature G pH value in flow cell H Rel...

Страница 56: ...that might need to be modified during daily routine Normally password protected and used by the process operator Subset of menu 5 Installation but process related Menu Installation 5 For initial inst...

Страница 57: ...s the new parameter is set 5 1 2 Sensors Sensor type FOME Temperature NT5K Standards Disinf Free chlorine 4 4 1 Logger Log interval 30 min Clear logger no 4 1 3 Logger Clear logger no Log interval 30m...

Страница 58: ...ndard bottle well and fill it with grab sample Do not use closed bottles 2 Screw the grab sample bottle into the standard bottle holder and swing it upwards This stops the sample flow from the constan...

Страница 59: ...4 After the grab sample bottle is empty swing the bottle holder down and unscrew the grab sample bottle 5 Press the key again HOLD appears in the display cal delay After the delay time is over the ins...

Страница 60: ...eplace it if necessary Check level of KCl reservoir If necessary replace the bottle Etch the sodium electrode using SWAN s etching kit Perform a two point calibration Perform a pH measurement and corr...

Страница 61: ...ntenance 1 Stop sample flow 2 Shut off power of the instrument 3 Disconnect the reagent bottle and close it 4 If the automatic regeneration module is installed swing down unscrew and close the regener...

Страница 62: ...e electrical connectors stay absolutely clean and dry Remove the sodium electrode 1 Unscrew and remove the sensor plug A Prevent the connector from getting wet 2 Put on the connector cap B onto the se...

Страница 63: ...before use and mark the date of mixing Note Once the fluoride salt is dissolved the life time of the solution is limited to 6 months Cleaning and etching 1 Remove adhered iron deposits by wiping the e...

Страница 64: ...e bottle was punctured do not spill KCl 5 Remove the reference electrode from the flow cell Cleaning 1 Remove any iron deposits with a soft paper tissue 2 Slip the ring shaped sleeve F upwards with a...

Страница 65: ...he union screw C fr om the threaded hole 4 Remove the pH electrode together with the union screw C washer D and O ring E from the measuring cell 5 If necessary wipe the pH sensor shaft and the green t...

Страница 66: ...ple Stream Option Disassemble the solenoid valve 1 Switch off the instrument according to instructions in Stop of Op eration for Maintenance p 61 2 Loosen the nut A 3 Remove the solenoid coil B from t...

Страница 67: ...r Note The O rings inside the valve body may stick on the flow cell and fall down if the valve body is removed 5 Remove the valve body from the flow cell 6 Remove the white plate G with a screw driver...

Страница 68: ...on solution tube and the flow cell completely 5 Switch off the instrument Dismount 1 Remove the tubes from the defective valve 2 Unscrew knurled nut A 3 Disconnect the coil body B from the solenoid va...

Страница 69: ...Place washer 2 on membrane and push down carefully 9 Place long spring in solenoid 10 Screw on the solenoid support fingertight Mounting 1 Mount the solenoid valve to the automatic regeneration modul...

Страница 70: ...e fragile and scratch sensitive Possible damage of acrylic glass parts due to scrubbing materials Never use organic solvents or scrubbing materials to clean acrylic glass parts Use soft detergent and...

Страница 71: ...l D completely 4 Remove all sensors A 5 Remove all tube connections 6 Unscrew the 4 screws B of the flow cell cover C and remove it 7 Clean the flow cell with a soft brush Assemble the flow cell 1 Scr...

Страница 72: ...ubbing materi als Never use organic solvents or scrubbing materials to clean acrylic glass parts Use soft detergent and rinse well Eliminate lime deposits with a common household deliming agent in sta...

Страница 73: ...ck 9 Clean all acrylic parts with a soft brush bottle cleaner and soapy water Note Grease the O rings only with teflon paste or spray Assemble the flow cell 1 Replace all O rings C before assembling t...

Страница 74: ...ert the air filter B in the holder A on the right of the transmitter To replace the air filter proceed as follows 1 Pull out the polluted air filter of the holder A 2 Push the new air filter B into th...

Страница 75: ...centrations are programmed correctly See menu 5 1 5 p 99 Prepare two standards Prepare the 2 standards directly in the marked bottle using the stock solution The final concentration must correspond to...

Страница 76: ...meter to do the correction We recom mend using a Chematest series instrument with a pH electrode The handheld pH meter must be calibrated correctly Navigate to menu Maintenance Calibration Process pH...

Страница 77: ...hing kit only Electrode Offset The instrument guides you through the complete calibration process If you have completed the required action press Enter to proceed 6 Press Enter to save 3 1 2 4 Process...

Страница 78: ...s Enter for 2 point calibration or Exit if you only want to make a one point calibration Standard Sodium and wait Enter to continue 3 1 1 5 Mount electrode 2 3 minutes Standard Sodium onto holder and...

Страница 79: ...outputs are frozen If the delay time is 0 the outputs show the mea sured value During calibration delay HOLD is displayed 1 Screw standard bottle 2 onto the holder and swing bottle upwards 2 Wait unti...

Страница 80: ...11 1 Tube Numbering The NO port of the 3 way valve must be sealed with a blind plug Tube No from to 1 Air filter A Reagent G 2 Reagent G Flow cell block C 3 Solenoid valve NC F T piece H 4 Solenoid va...

Страница 81: ...ubble detector 6 Finally slide the other light protecting sleeve with the rounded side facing towards the flow cell cover over the reaction tube Note Extraneous light may distort the bubble measuremen...

Страница 82: ...e immediately show the label where possible 1 Carefully unscrew and remove the screw cover with tube holder from the DIPA bottle 2 Close the DIPA bottle with the original cover 3 Remove the EPDM seal...

Страница 83: ...and close it Rinse the tube with sample water using the Maintenance Regeneration Manual Operation menu item pH sensor 8 Unscrew and remove the connector from the pH sensor 9 Put the connector cap ont...

Страница 84: ...appear if correct measuring conditions are re covered i e Sample Flow low Such Errors are marked E0xx bold and orange Errors which indicate a hardware failure of the instrument Such Errors are marked...

Страница 85: ...eck programmed value 5 3 1 1 2 p 104 check if 2nd stream is installed correctly E006 Sodium 2 Alarm low check process check programmed value 5 3 1 1 2 p 104 check if 2nd stream is installed correctly...

Страница 86: ...in Installation Relay 5 3 2 and 5 3 3 p 106 E018 Reagent empty check DIPA bottle if necessary refill diisopropylamine E019 No Sample readjust the sample flow check if tubes are blocked E020 pH low che...

Страница 87: ...ssible Reasons No automatic regeneration is performed The automatic regeneration does not start if one of the following errors is present E010 Sample flow low E018 Reagent empty In addition the automa...

Страница 88: ...ng installation relay 1 relay 2 alarm relay When a fuse has blown find out the cause and fix it before replacing it with a new one Use tweezers or needle nosed pliers to remove the defective fuse Use...

Страница 89: ...gs can be modified Menu 3 Maintenance is for service Calibration simulation of outputs and set time date Please protect with password Menu 4 Operation is for the user allowing to set limits alarm valu...

Страница 90: ...ensors Sodium Sensor Current Value 2 2 2 2 1 Raw value 1 Cal History Number 2 2 1 5 1 2 2 1 5 Date Time Offset Slope pH Electrode Current Value 2 2 2 Raw value Cal History Number 2 2 2 5 1 2 2 2 5 Dat...

Страница 91: ...ave Enter 3 1 2 5 Regeneration Manual Operation Valve Open 3 2 3 2 1 3 2 1 1 Closed Parameters Start time 3 2 2 1 3 2 2 Calendar Monday to Sunday 3 2 2 2 5 1 2 2 1 to 5 1 2 2 7 Duration 3 2 2 3 Delay...

Страница 92: ...teresis 4 2 1 3 x Delay 4 2 1 3 x Relay 1 and 2 Setpoint 4 2 x x 4 2 2 and 4 2 3 Hysteresis 4 2 x x Delay 4 2 x x Input Active 4 2 4 1 4 2 4 Signal Outputs 4 2 4 2 Output Control 4 2 4 3 Fault 4 2 4 4...

Страница 93: ...m Low 5 3 1 5 x Case Temp high 5 3 1 6 Case Temp low 5 3 1 7 Relay 1 and 2 Function 5 3 2 1 5 3 3 1 5 3 2 5 3 3 Parameter 5 3 2 x 5 3 3 x Setpoint 5 3 2 x 5 3 3 x Hysteresis 5 3 2 x 5 3 3 x Delay 5 3...

Страница 94: ...only be viewed not modified 2 1 Identification Designation View the designation of instrument Version Firmware of instrument e g V6 22 06 21 2 1 3 Factory Test Test date of the instrument motherboard...

Страница 95: ...s current status of all in and outputs 2 4 1 2 4 2 2 5 Interface Only available if optional interface is installed Shows the programmed communication settings 3 Maintenance 3 1 Calibration In this men...

Страница 96: ...2 4 Delay During regeneration plus delay time parameters 3 2 2 5 p 96 and 3 2 2 6 p 96 apply Range 0 6 000 Sec 3 2 2 5 Signal Outputs Select the operation mode of the signal outputs during regenerati...

Страница 97: ...Adjust date and time 4 Operation 4 1 Sensors 4 1 1 Filter Time Constant Used to damp noisy signals The higher the fil ter time constant the slower the system reacts to changes of the measured value Ra...

Страница 98: ...interface is in stalled The logger can save approx 1500 data records Records consists of Date time alarms measured value pH temperature flow 4 3 1 Log Interval Select a convenient log interval Consult...

Страница 99: ...0 00 ppb 20 ppm 5 1 5 2 Standard 2 Solution 2 high concentration Select concentration at least 10 times higher than solution 1 Range 0 00 ppb 20 ppm 5 2 Signal Outputs 5 2 1and 5 2 2 Signal Output 1 a...

Страница 100: ...linear or logarithmic scale In addition the midpoint for the bilinear scale Parameter Sodium 1 5 2 1 40 10 Range low 0 ppb 20 ppm 5 2 1 40 20 Range high 0 ppb 20 ppm Parameter Sodium 2 5 2 1 40 11 Ra...

Страница 101: ...y the P Band In the steady state the setpoint will never be reached The deviation is called steady state error Parameters setpoint P Band PI controller The combination of a P controller with an I cont...

Страница 102: ...fined process value Measured value or flow P Band Range below upwards control or above downwards control the set point within the dosing intensity is reduced from 100 to 0 to reach the set point witho...

Страница 103: ...will be sudden ly reached by a D controller Range 0 9 000 sec 5 2 1 43 5 Control timeout If a controller action dosing intensity is constantly over 90 during a defined period of time and the process v...

Страница 104: ...ontrol out puts are held in the operating mode Range 0 00 28 800 sec 5 3 1 3 Alarm pH Define the measuring value which should issue an alarm high respectively low 5 3 1 3 1 Alarm High If the measured...

Страница 105: ...m Low If the measuring values falls below the programmed val ue E010 will be issued Range 5 1000 B s 5 3 1 5 Sample Temperature Define the measuring value which should is sue an alarm high respectivel...

Страница 106: ...hen enter the necessary data depending on the selected function When the relays are used as upper or lower limit switches program the following 5 3 2 20 Parameter select a process value 5 3 2 300 Setp...

Страница 107: ...cy Motor valve Examples of metering devices that are driven time proportional are solenoid valves peristaltic pumps Dosing is controlled by the operating time 5 3 2 32 20 Cycle time duration of one co...

Страница 108: ...the programmed time scheme 5 3 2 24 Mode Operating mode interval daily weekly 5 3 2 340 Interval The interval can be programmed within a range of 1 1 440 min 5 3 2 44 Run Time Enter the time the relay...

Страница 109: ...5 3 2 6 Signal Outputs see Interval 5 3 2 7 Output Control see Interval The relay contact can be activated at one or several days of a week The daily starting time is valid for all days 5 3 2 342 Cale...

Страница 110: ...Sample Stream Option via an external device 5 3 4 1 Active Define when the input should be active The measurement is interrupted during the time the input is active 5 3 4 2 Signal Outputs Select the...

Страница 111: ...lt All other values are kept in memory In parts Communication parameters are kept in memory All other values are set back to default values Completely Sets back all values including communication para...

Страница 112: ...number 5 4 6 Line Break Detection Define if message E028 should be issued in case of a line break on signal output 1 or 2 Choose between Yes or No 5 5 Interface Select one of the following communicati...

Страница 113: ...oading at www swan ch Catalogue No Product name A 87 729 010A Etching Kit A Catalogue No Product name A 87 729 010B Etching Kit B Catalogue No Product name A 85 141 400 Sodium Standard Solution 1 000...

Страница 114: ...on Signal Output same as in Installation Relay 1 and 2 same as in Installation Input same as in Installation Logger Logger Interval 30 min Clear Logger no Installation Sensor Sensor type Sodium Temper...

Страница 115: ...C Case temp low 0 C Relay 1 and 2 Function Limit upper Parameter Relay1 and 2 Soduim 1 Setpoint Relay1 and 2 1 00 ppm Hysteresis 10 ppb Delay 30 s If Function Control upw or dnw Parameter Sodium 1 an...

Страница 116: ...Timeout 0 min Settings Actuator Time proportional Cycle time 60 s Response time 10 s Settings Actuator Motor valve Run time 60 s Neutral zone 5 If Function Timer Mode Interval Interval 1 min Mode dai...

Страница 117: ...e 95 Identification 94 Input 43 Instrument Setup 21 Interface 93 95 HART 48 Modbus 47 Profibus 47 USB 48 L Logger 114 Longer Stop of Operation 83 M Maintenance Reference electrode 64 Sodium electrode...

Страница 118: ...AMI Sodium P Index A 96 250 211 171121 118 Sensors 92 94 114 Signal Outputs 46 92 114 Simulation 97 Software Structure 56 T Temperature 95 Terminals 41 43 44 47 U USB Interface 48 W Wire 39...

Страница 119: ...A 96 250 211 171121 AMI Sodium P Notes 119 13 Notes...

Страница 120: ...ast e t W a e r E f u e n t s P o o l S a n i t a r y Swan Analytical Instruments CH 8340 Hinwil www swan ch swan swan ch Swan Products Analytical Instruments for Swan is represented worldwide by sub...

Отзывы: