AMI Sodium P
Program List and Explanations
104
5.3.1.1/2
Alarm Sodium 1 and 2
5.3.1.x.x
Alarm High:
If the measured value rises above the alarm high value,
the alarm relay is activated and E001 is displayed in the message
list.
Range: 0.00–20.00 ppm
5.3.1.x.x
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E002 is displayed in the message
list.
Range: 0.00–20.00 ppm
5.3.1.x.x
Hysteresis:
Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluctu
-
ates around the alarm value.
Range. 0.00–20.00 ppm
5.3.1.x.x
Delay:
During run time plus the delay time the signal and control out
-
puts are held in the operating mode.
Range: 0.00–28‘800 sec
5.3.1.3
Alarm pH:
Define the measuring value, which should issue an alarm
high respectively low.
5.3.1.3.1
Alarm High:
If the measured value rises above the alarm high value,
the alarm relay is activated and E003 is displayed in the message
list.
Range: 0–14.00 pH
5.3.1.3.25
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E004 is displayed in the message list
Range: 0–14.00 pH
5.3.1.3.35
Hysteresis:
Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluctu
-
ates around the alarm value
Range: 0–14.00 pH
5.3.1.3.45
Delay:
During run time plus the delay time the signal and control out
-
puts are held in the operating mode
Range: 0–28‘800 Sec
5.3.1.4
Sample Flow:
Define at which sample flow a flow alarm should be
issued.