AMI Oxytrace QED
Program List and Explanations
A-96.250.531 / 300617
81
5.3.2 and 5.3.3
Relay 1 and 2:
The function of relay contacts 1 or 2 are defined by
the user
NOTICE:
The navigation in the menu <Relay 1> and <Relay 2>
is equal. For reason of simplicity only the menu numbers of
Relay 1 are used in the following.
1
First select the functions as:
- Limit upper/lower,
- Control upwards/downwards,
- Timer
- Fieldbus,
2
Then enter the necessary data depending on the selected func-
tion. The same values may also be entered in menu
When the relays are used as upper or lower limit switches, program
the following:
5.3.2.20
Parameter:
choose one of the following process values
Oxygen
Temperature
Sample Flow
Saturation
5.3.2.300
Setpoint
: If the measured value rises above respectively falls below
the set-point, the relay is activated.
5.3.2.400
Hysteresis:
within the hysteresis range, the relay does not switch.
This prevents damage of relay contacts when the measured value
fluctuates around the alarm value.
5.3.2.1
Function = Limit upper/lower:
Parameter
Range
Oxygen
0.00 ppb –20.00 ppm
Temperature
-30 to + 130 °C
Sample flow
0–50 l/h
Saturation
0–200%
Parameter
Range
Oxygen
0.00 ppb –20.00 ppm
Temperature
0–100 °C
Sample flow
0–50 l/h
Saturation
0–200%
Содержание AMI Oxytrace
Страница 1: ...AMI Oxytrace AMI Oxytrace QED Version 6 20 and higher A 96 250 531 300617 Operator s Manual...
Страница 5: ...AMI Oxytrace QED A 96 250 531 300617 3 11 Default Values 90 12 Index 93 13 Notes 94...
Страница 96: ...94 A 96 250 531 300617 AMI Oxytrace QED Notes 13 Notes...
Страница 97: ...AMI Oxytrace QED Notes A 96 250 531 300617 95...