42
A-96.250.531 / 300617
AMI Oxytrace QED
Maintenance
6.3.
Maintenance of the Oxygen Sensor
WARNING
Etching liquid
The electrolyte is alkaline and caustic. It contains less than 1%
of potassium hydroxide.
Do not ingest. Wear protective goggles and gloves during
handling. Avoid contact with clothes.
In case of accidental contact with the eyes, wash immediately
with clear water and contact a physician. Show him the label
of the bottle or this section of the manual.
Short contact with skin is harmless, nevertheless wash with
lots of water.
6.3.1
Electrolyte exchange
An electrolyte exchange is indicated in the maintenance list as
soon as the remaining amount is below 10%.
A
B
C
Fixing sleeve
Sensor
Flow cell
1
Unscrew the fixing sleeve [A].
2
Remove the sensor from the flow
cell.
NOTICE:
Do not pull out the sensor
with force!
To remove the sensor from the flow
cell, open the flow regulating valve
during removing the sensor. This
prevents the formation of a vacuum,
and the sensor can easily be
removed from the flow cell.
A
C
B