AMI Oxytrace QED
Error List
A-96.250.531 / 300617
61
E013
Case Temp. high
– check case/environment temperature
– check programmed value,
E014
Case Temp. low
– check case/environment temperature
– check programmed value,
E017
Control Timeout
– check control device or programming in
Installation, Relay contact, Relay 1/2
E018
Faraday Efficiency
– clean Faraday electrode, see
Maintenance of the Faraday electrode,
– perform an air calibration, see
– perform maintenance of the oxygen
E019
Quality Assurance
– Perform QA Procedure using a
reference instrument, e.g. AMI Inspector
E024
Input active
– See If Fault Yes is programmed in Menu
E026
IC LM75
– call service
E028
Signal output open
– check wiring on signal outputs 1 and 2
E030
EEProm Frontend
– call service
E031
Calibration Recout
– call service
Error
Description
Corrective action