90
A-96.250.531 / 300617
AMI Oxytrace QED
Default Values
11.
Default Values
Operation:
Sensors:
Filter Time Const.:..................................................................10 Sec
Hold after Cal.:.....................................................................300 Sec
Faraday Parameter:
Mode: ................................................................................ Interval
Interval:..............................................................................3 hours
Delay .................................................................................. 60 sec
Signal Outputs........................................................................ hold
Output/Control ....................................................................... hold
Alarm Relay
...................................................................... same as in Installation
Relay 1/2
...................................................................... same as in Installation
Input
...................................................................... same as in Installation
Logger:
Logger Interval:.............................................................. 30 Minutes
Clear Logger: ................................................................................no
Installation:
Sensors
Miscellaneous:
Flow:.................................................................................. Q-Flow
O2 Offset: ..........................................................................0.0 ppb
Quality Assurance:
Level:.................................................................................... 0: Off
Signal Output 1
Parameter: ........................................................................... Oxygen
Current loop: .....................................................................4 –20 mA
Function: .................................................................................. linear
Scaling: Range low: ............................................................0.00 ppb
Scaling: Range high:........................................................10.00 ppm
Signal Output 2
Parameter: ....................................................................Temperature
Current loop: .....................................................................4 –20 mA
Function: .................................................................................. linear
Scaling: Range low: ................................................................0.0 °C
Scaling: Range high:.............................................................50.0 °C
Alarm Relay:
Alarm oxygen; Alarm high:...............................................10.00 ppm
Alarm oxygen; Alarm low: ...................................................0.00 ppb
Alarm oxygen; Hysteresis: ...................................................100 ppb
Alarm oxygen; Delay:...............................................................5 Sec
Sample Flow, Flow Alarm: .......................................................... yes
Sample Flow, Alarm high: .................................................... 25.0 l/h
Sample Flow, Alarm low:........................................................ 8.0 l/h
Содержание AMI Oxytrace
Страница 1: ...AMI Oxytrace AMI Oxytrace QED Version 6 20 and higher A 96 250 531 300617 Operator s Manual...
Страница 5: ...AMI Oxytrace QED A 96 250 531 300617 3 11 Default Values 90 12 Index 93 13 Notes 94...
Страница 96: ...94 A 96 250 531 300617 AMI Oxytrace QED Notes 13 Notes...
Страница 97: ...AMI Oxytrace QED Notes A 96 250 531 300617 95...