88
A-96.250.531 / 300617
AMI Oxytrace QED
Program List and Explanations
5.5 Interface
Select one of the following communication protocols. Depending on
your selection, different parameters must be defined.
5.5.1
Protocol:
Profibus
5.5.20
Device address:
Range: 0–126
5.5.30
ID-Nr.:
Range: Analyzer; Manufacturer; Multivariable
5.5.40
Local operation:
Range: Enabled, Disabled
5.5.1
Protocol:
Modbus RTU
5.5.21
Device address:
Range: 0–126
5.5.31
Baud Rate:
Range: 1200–115 200 Baud
5.5.41
Parity:
Range: none, even, odd
5.5.1
Protocol:
USB-Stick
Only visible if an USB interface is installed. No further settings are
possible.
5.5.1
Protocol
:
HART
5.5.24
Device address:
Range: 0–63