AMI Oxytrace QED
Installation
CAUTION
Risk of damage of the relays in the AMI Transmitter due to
heavy inductive load.
Heavy inductive or directly controlled loads (solenoid valves,
dosing pumps) may destroy the relay contacts.
To switch inductive loads > 0.1 A use an AMI relay box avail-
able as an option or suitable external power relays.
Inductive load
Small inductive loads (max 0.1A) as for example the coil of a power
relay can be switched directly. To avoid noise voltage in the
AMI Transmitter it is mandatory to connect a snubber circuit in par-
allel to the load.
A snubber circuit is not necessary if an AMI relaybox is used.
Resistive load
Resistive loads (max. 1A) and control signals for PLC, impulse
pumps and so on can be connected without further measures
Actuators
Actuators, like motor valves, are using both relays: One relay con-
tact is used for opening, the other for closing the valve, i.e. with the
2 relay contacts available, only one motor valve can be controlled.
Motors with loads bigger than 0.1A must be controlled via external
power relays or an AMI relay box.
A
B
C
D
E
AC or DC power supply
AMI Transmitter
External power relay
Snubber
Power relay coil
A
B
C
D
E
A
B
C
AMI Transmitter
PLC or controlled pulse pump
Logic
A
B
C
A
B
C
AC or DC power supply
AMI Transmitter
Actuator
M
A
B
C
Содержание AMI Oxytrace
Страница 1: ...AMI Oxytrace AMI Oxytrace QED Version 6 20 and higher A 96 250 531 300617 Operator s Manual...
Страница 5: ...AMI Oxytrace QED A 96 250 531 300617 3 11 Default Values 90 12 Index 93 13 Notes 94...
Страница 96: ...94 A 96 250 531 300617 AMI Oxytrace QED Notes 13 Notes...
Страница 97: ...AMI Oxytrace QED Notes A 96 250 531 300617 95...