A-96.250.531 / 300617
AMI Oxytrace QED
Program List and Explanations
5.3.1.2.2
Alarm High:
If the measuring values rises above the programmed
value E009 will be issued.
Range: 12–50 l/h
5.3.1.2.35
Alarm Low:
If the measuring values falls below the programmed
value E010 will be issued.
Range: 8–11 l/h
5.3.1.3
Sample Temp.:
Define at which sample temperature an alarm
should be issued.
5.3.1.3.1
Alarm High:
If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E007 is issued.
Range: 30–100 °C
5.3.1.3.25
Alarm Low:
If the measured value rises above the alarm high value,
the alarm relay is activated and E008 is issued.
Range: -10 to + 20 °C
5.3.1.4
Alarm Saturation
5.3.1.4.1
Alarm High:
If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E003, is displayed in the mes-
sage list.
Range: 0.00 –200 %
5.3.1.4.25
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E004 is displayed in the message
list.
Range: 0.00 –200 %
5.3.1.4.35
Hysteresis:
Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluc-
tuates around the alarm value.
Range. 0.00 –200 %
5.3.1.4.45
Delay:
Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 Sec
5.3.1.5
Case Temp.
5.3.1.5.1
Case Temp. high:
Set the alarm high value for temperature of elec-
tronics housing. If the value rises above the programmed value
E013 is issued.
Range: 30–75 °C
5.3.1.5.2
Case Temp. low:
Set the alarm low value for temperature of elec-
tronics housing. If the value falls below the programmed value
E014 is issued.
Range: -10 to + 20 °C
Содержание AMI Oxytrace
Страница 1: ...AMI Oxytrace AMI Oxytrace QED Version 6 20 and higher A 96 250 531 300617 Operator s Manual...
Страница 5: ...AMI Oxytrace QED A 96 250 531 300617 3 11 Default Values 90 12 Index 93 13 Notes 94...
Страница 96: ...94 A 96 250 531 300617 AMI Oxytrace QED Notes 13 Notes...
Страница 97: ...AMI Oxytrace QED Notes A 96 250 531 300617 95...