72
A-96.250.531 / 300617
AMI Oxytrace QED
Program List and Explanations
3.3 Simulation
In this menu the following relays and signal outputs can be tested:
Alarm relay
Relay 1 and 2
Signal output 1 and 2
Signal output 3 (if option is installed)
Select a relay or signal output with the [
] or [
] keys,
press the [Enter]> key to confirm. Then change the value with the
[
] or [
] keys. After confirming the setting with the [Enter]
key, the value is simulated by the relay/signal output.
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated val-
ues will be reset.
3.4 Set Time
Adjust date and time.
3.5 Quality Assurance
Performs a Quality Assurance according to your settings. Follow
the commands on the screen. Detailed explanation see
surance of the Instrument, p. 51
.
Alarm Relay:
Active or inactive.
Relay 1 and 2:
Active or inactive.
Input:
Open or closed.
Signal Output
1 and 2
:
Current in mA
Signal Output 3:
Current in mA (if option is installed)