22
A-96.250.531 / 300617
AMI Oxytrace QED
Installation
3.2.
Mounting of Instrument Panel
The first part of this chapter describes the preparing and placing of
the system for use.
The instrument must only be installed by trained personnel.
Mount the instrument in vertical position.
For ease of operation mount it so that the display is at eye
level.
For the installation a kit containing the following installation
material is available:
– 4 Screws 8x60 mm
– 4 Dowels
– 4 Washers 8.4/24 mm
Mounting re-
quirements
The instrument is only intended for indoor installation.
Dimensions (Oxytrace and Oxytrace QED), p.
Turn-on
power
Switch on power.
First, the analyzer performs a self test, displays the firmware ver-
sion and then starts normal operation
Run-in Period
Calibration
Leave the sensor at the air. The sensor has to run-in for at least
30 min, better 1 h. During this time it has to be at air with a dry and
clean membrane and has to be connected to the powered instru-
menty. Calibrate the sensor after this time, see
.
and mount it into the flow cell afterwards.
Open
Sample Flow
Open the flow regulating valve to allow sample flow into the flow
cell and into the waste. Sample must always overflow into waste
Programming
Program all parameters for external devices (interface, etc.). Set
all parameters for instrument operation (limits, alarms).
Содержание AMI Oxytrace
Страница 1: ...AMI Oxytrace AMI Oxytrace QED Version 6 20 and higher A 96 250 531 300617 Operator s Manual...
Страница 5: ...AMI Oxytrace QED A 96 250 531 300617 3 11 Default Values 90 12 Index 93 13 Notes 94...
Страница 96: ...94 A 96 250 531 300617 AMI Oxytrace QED Notes 13 Notes...
Страница 97: ...AMI Oxytrace QED Notes A 96 250 531 300617 95...