50
A-96.250.531 / 300617
AMI Oxytrace QED
Maintenance
6.7.
Faraday Verification
The Faraday verification works only for oxygen concentrations be-
low 200 ppb. If automatic Faraday verification is enabled, a periodic
check of the system is performed. A manual verification can be
started for test purposes.
Automatic
Verification
Per default the instrument performs an automatic faraday verifica-
tion every 3 hours. To change the settings for automatic verification
go to menu Faraday Parameter- Timer mode, see menu
for details.
Possible settings are:
off
interval
daily
weekly
Manual
Verification
To start a manual verification:
1
Navigate to menu 3.2.2 <Maintenance>/<Service>/<Faraday
Verification>.
Results are saved in the Verification history menu 2.2.1.5
2
Press [Enter] to start the Faraday
Verification.
The verification then starts
immediately.
3
Press [Enter] to confirm the Fara-
day Verification.
3.3
Progress
Faraday Verification
<Enter> to stop
Current Value
1.62
ppb
Faraday Conc.
12.85
ppb
3.3
Progress
Faraday Verification
Done
Efficiency
91.5 %
Faraday Conc.
12.85
ppb