5.4 Miscellaneous
5.4.1
Language:
Set the desired language.
Available settings: German /English/French/Spanish
5.4.2
Set defaults:
Reset the instrument to factory default values in three
different ways:
Calibration:
Sets calibration values back to default. All other
values are kept in memory.
In parts:
Communication parameters are kept in memory. All
other values are set back to default values.
Completely:
Sets back all values including communication
parameters.
5.4.3
Load Firmware:
Firmware updates should be done by instructed
service personnel only.
5.4.4
Password:
Select a password different from 0000 to prevent unau-
thorized access to the menus “Messages”, “Maintenance”, “Opera-
tion” and “Installation”.
Each menu may be protected by a
different
password.
If you forgot the passwords, contact the closest SWAN representa-
tive.
5.4.5
Sample ID:
Identify the process value with any meaning full text,
such as KKS number.
5.4.6
Line Break Detection:
If activated, error message E028 is shown in
case of line break on signal outputs 1 and 2.