74
A-96.250.531 / 300617
AMI Oxytrace QED
Program List and Explanations
4.2 Relay Contacts
4.3 Logger
The instrument is equipped with an internal logger. The data can be
copied to the USB stick installed in the transmitter.
The logger can save approx. 1500 data records. The Records con-
sists of: Date, time, alarms, measuring values, raw values, case
temperature, flow.
4.3.1
Log Interval:
Select a convenient log interval. Consult the table be-
low to estimate the max logging time. When the logging buffer is
full, the oldest data record is erased to make room for the newest
one (circular buffer).
Range: 1 Second to 1 hour
4.3.2
Clear Logger:
If confirmed with
yes
, the complete logger data is de-
leted. A new data series is started.
5 Installation
5.1 Sensors
5.1.1
Miscellaneous
5.1.1.1
Flow
:
If a flow cell without flow measurement (e.g. B-Flow) is used,
choose none. With flow measurement select Q-Flow
5.1.1.2
O2 Offset
: Manual, small correction of the offset.
Range -5 to 5 ppb
5.1.2
Quality Assurance:
Switch the Quality Assurance on or off.
5.1.2.1
Level
: Select quality level:
Level 0: Off
Quality assurance procedure switched off. Any additional QA
menus are hidden.
Level 1: Trend
Level 2: Standard
Level 3: Crucial
Level 4: User
Edit user specific limits in menu 5.1.2.2
Interval
1 s
5 s
1 min
5 min
10 min
30 min
1 h
Time
25 min
2 h
25 h
5 d
10 d
31 d
62 d
Содержание AMI Oxytrace
Страница 1: ...AMI Oxytrace AMI Oxytrace QED Version 6 20 and higher A 96 250 531 300617 Operator s Manual...
Страница 5: ...AMI Oxytrace QED A 96 250 531 300617 3 11 Default Values 90 12 Index 93 13 Notes 94...
Страница 96: ...94 A 96 250 531 300617 AMI Oxytrace QED Notes 13 Notes...
Страница 97: ...AMI Oxytrace QED Notes A 96 250 531 300617 95...