AMI Oxytrace QED
Installation
A-96.250.531 / 300617
21
3.
Installation
The first part of this chapter describes the preparing and placing of
the system in position for use. The second part describes the wiring
and the properties of the signal outputs and relay.
Restrictions
on use
If exposed to weather, install a cover to protect the instrument from
insolation and precipitation. See
for tempera-
ture limitations.
3.1.
Installation Check List
Check
Instrument’s specification must conform to the National Electrical
Code, all state and local codes, and all plant codes and standards
for electrical equipment.
On site
requirements
100–240 VAC (
10%), 50/60 Hz (
5%) or 24 VDC (±10%),
isolated power outlet with ground connection and 30 VA
Sample line with 8–25 l/h and 0.2–3 bar.
Waste line with pressure free drain.
Installation
Mount the instrument in vertical position. Display should be at eye
level.
Connect the sample and waste line, see
Electrical
Connections
Connect all external devices like limit switches, current loops and
pumps, see
.
Connect the power cord, do not switch on power until all external
devices are connected.
Swansensor
Oxytrace G
The Swansensor Oxytrace G is delivered with prefilled electrolyte
chamber. A transport protection cap filled with water keeps the
sensor wet during transport and storage. Before installation,
remove the transport protection cap and clean the sensor sur-
faces with water. Afterwards install the oxygen sensor and con-
nect the cable.
Install the Swansensor Oxytrace G, p. 24
.
Содержание AMI Oxytrace
Страница 1: ...AMI Oxytrace AMI Oxytrace QED Version 6 20 and higher A 96 250 531 300617 Operator s Manual...
Страница 5: ...AMI Oxytrace QED A 96 250 531 300617 3 11 Default Values 90 12 Index 93 13 Notes 94...
Страница 96: ...94 A 96 250 531 300617 AMI Oxytrace QED Notes 13 Notes...
Страница 97: ...AMI Oxytrace QED Notes A 96 250 531 300617 95...