A-96.250.511 / 091120
AMI Turbiwell
Maintenance
67
Verification
For possible error messages see
.
1
Navigate to menu <Maintenance>/
<Verification> and press [Enter].
2
Select SWAN Verikit and press
[Enter].
3
Choose a Verikit from the list.
4
Press [Enter].
5
Stop sample flow.
6
Open the measuring chamber.
7
Open the drain valve if the instru
-
ment is not equipped with an auto
-
matic drain valve.
8
Wait until the measuring chamber is
empty.
9
Close the drain valve.
10
Install the verification kit (see
Installation of the Verikit, p. 68
).
11
Close the measuring chamber.
12
Wait until the verification is finished.
13
Open the measuring chamber.
14
Remove the verification kit.
15
Close measuring chamber.
16
Start sample flow.
The Verification results are stored
in the Verification history, see
.
3.1.2
Verification
Other method
SWAN Verikit
3.1.1
Verification
<Enter> to continue
VERIKIT #
______
VERIKIT #
001
002
003
3.1.5
- Close inlet valve
- Empty meas. chamber
Verification
- Open meas. chamber
- Drainage valve open?
<Enter> to continue
3.1.5
- Drainage valve closed?
Verification
- Mount test unit
- Close meas. chamber
<Enter> to continue
3.1.5
Actual value
21.7 FNU
Verification
Reference value
21.6 FNU
Deviation
0.1%
<Enter> to save
3.1.5
- Open meas. chamber
- Open inlet valve
Verification
- Remove Test unit
- Close meas. chamber
<Enter> to finish
Содержание AMI Turbiwell
Страница 1: ...Operator s Manual Firmware V6 23 and higher AMI Turbiwell A 96 250 511 091120 ...
Страница 35: ...A 96 250 511 091120 AMI Turbiwell Installation 35 Degasser G B C Degasser sample inlet Drain Waste B G C ...
Страница 108: ...A 96 250 511 091120 AMI Turbiwell Notes 108 12 Notes ...
Страница 109: ...A 96 250 511 091120 AMI Turbiwell Notes 109 ...