Instrument Setup
50
Process
calibration
The following must be considered before starting calibration:
Drain the measuring chamber and clean it if necessary.
For the calibration, oil-free process water and oil is needed.
The calibration solutions are prepared in a vessel with a vol
-
ume of approx.10 liters.
Homogenize the solution with a circulating pump or a motor
-
ized stirrer before and during the calibration.
Connect the outlet of the vessel with the sample inlet of the
AMI Turbiwell.
Adjust the sample flow to approx. 20 liters per hour by a control
valve.
The signal must be stable while the sample is flowing. To check
the stability, navigate to menu 2.2.1 <Diagnostic>/<Sensors>/
<Turbidity> and observe the raw signal.
To start with calibration navigate to menu 3.1, <Maintenance>/
<Process Cal. ppm>. There you can choose between
Detect Zero:
The zero value determination is carried out with oil-free pro
-
cess water.
Specify slope:
A sample with a known oil concentration is used to determine
the slope. The entered process value is used as the setpoint of
the calibration sample.
Содержание AMI Turbiwell
Страница 1: ...Operator s Manual Firmware V6 23 and higher AMI Turbiwell A 96 250 511 091120 ...
Страница 35: ...A 96 250 511 091120 AMI Turbiwell Installation 35 Degasser G B C Degasser sample inlet Drain Waste B G C ...
Страница 108: ...A 96 250 511 091120 AMI Turbiwell Notes 108 12 Notes ...
Страница 109: ...A 96 250 511 091120 AMI Turbiwell Notes 109 ...