A-96.250.511 / 091120
AMI Turbiwell
Installation
28
6
Turn the elbow hose nozzle [E] upwards clockwise.
7
Install the locking plate [I]
8
Tighten the screw [J].
9
Screw the deltaT sensor [B] to the panel in vertical position.
10
Connect the tube supplied with the installation kit to the sample
outlet [C] of the deltaT sensor and to the sample inlet (elbow
hose nozzle [E]) of the constant head.
11
Connect the sample inlet tube to sample inlet [A] of the deltaT
sensor.
12
Push the locking pin upward and swivel the measuring chamber
back.
13
Make sure that the measuring chamber rests in locked position.
Electrical
Connection
WARNING
Electrical shock hazard
Before opening the AMI Transmitter switch power off.
14
Use one of the PG7 cable glands to feed the cable of the sensor
into the AMI transmitter housing.
15
Connect the cable to the terminals according to the