A-96.250.511 / 091120
AMI Turbiwell
Default Values
104
10. Default Values
Operation
Sensors
Filter Time Const.: ........................................................................30 s
Hold after Cal.:............................................................................300 s
Alarm Relay
.........................................................................same as in Installation
Relay 1 & 2
.........................................................................same as in Installation
Input
.........................................................................same as in Installation
Logger
Logger Interval:........................................................................ 30 min
Clear Logger:.................................................................................. no
Installation
Sensor
Dimension:...................................................................................FNU
Flow; Flow measurement: ..........................................................None
Signal Output
1 & 2
Parameter:.......................................................................Meas. value
Current loop:........................................................................4 –20 mA
Function:.....................................................................................linear
Scaling: Range low:......................................................0.0 FNU/NTU
Scaling: Range high: .................................................. 10.0 FNU/NTU
Alarm Relay
Alarm high: ..................................................................200 FNU/NTU
Alarm low:................................................................. 0.000 FNU/NTU
Hysteresis:..................................................................... 10 FNU/NTU
Delay: .............................................................................................5 s
Sample Flow: Flow Alarm:.............................................................yes
Sample Flow: Alarm High: ......................................................40.0 l/h
Sample Flow: Alarm Low:......................................................... 6.0 l/h
Case temp. high: ....................................................................... 65 °C
Case temp. low:........................................................................... 0 °C
Relay 1 & 2
Function:.............................................................................limit upper
Parameter:.......................................................................Meas. value
Setpoint: ......................................................................100 FNU/NTU
Hysteresis:..................................................................5.00 FNU/NTU
Delay: .............................................................................................5 s
If Function = Control upw. or dnw:
Parameter: Meas. value
Settings: Actuator: Frequency
Settings: Pulse Frequency: .................................................120/min
Settings: Control Parameters: Setpoint:...................100 FNU/NTU
Settings: Control Parameters: P-band: ................... 5.00 FNU/NTU
Settings: Control Parameters: Reset time:..................................0 s
Содержание AMI Turbiwell
Страница 1: ...Operator s Manual Firmware V6 23 and higher AMI Turbiwell A 96 250 511 091120 ...
Страница 35: ...A 96 250 511 091120 AMI Turbiwell Installation 35 Degasser G B C Degasser sample inlet Drain Waste B G C ...
Страница 108: ...A 96 250 511 091120 AMI Turbiwell Notes 108 12 Notes ...
Страница 109: ...A 96 250 511 091120 AMI Turbiwell Notes 109 ...