![Swan Analytical Instruments AMI Turbiwell Скачать руководство пользователя страница 99](http://html1.mh-extra.com/html/swan-analytical-instruments/ami-turbiwell/ami-turbiwell_operators-manual_1411373099.webp)
A-96.250.511 / 091120
AMI Turbiwell
Program List and Explanations
99
5.3.2.32.20
Cycle time:
duration of one control cycle (on/off change).
Range: 0–600 sec.
5.3.2.32.30
Response time:
Minimal time the metering device needs to react.
Range: 0–240 sec.
5.3.2.32.4
Control Parameters:
Range for each Parameter same as
Examples of metering devices that are pulse frequency driven are
the classic membrane pumps with a potential free triggering input.
Dosing is controlled by the repetition speed of dosing shots.
5.3.2.32.21
Pulse frequency:
Max. pulses per minute the device is able to re
-
spond to. Range: 20–300/min.
5.3.2.32.31
Control Parameters:
Range for each Parameter same as
Dosing is controlled by the position of a motor driven mixing valve.
5.3.2.32.22
Run time:
Time needed to open a completely closed valve
Range: 5–300 Sec.
5.3.2.32.32
Neutral zone:
Minimal response time in % of the runtime. If the re
-
quested dosing output is smaller than the response time, no change
will take place.
Range: 1–20 %
5.3.2.32.4
Control Parameters:
Range for each Parameter same as
The relay will be activated repetitively depending on the programmed
time scheme.
5.3.2.24
Mode:
Operating mode (interval, daily, weekly)
5.3.2.340
Interval:
The interval can be programmed within a range
of 1–1’440 min.
5.3.2.44
Run Time
: Enter the time the relay stays active.
Range: 5–32’400 Sec.
5.3.2.54
Delay
: during run time plus the delay time the signal and control out
-
puts are held in the operating mode programmed below.
Range: 0–6’000 Sec.
Actuator = Frequency
Actuator = Motor valve
5.3.2.1
Function = Timer:
5.3.2.24
Interval
Содержание AMI Turbiwell
Страница 1: ...Operator s Manual Firmware V6 23 and higher AMI Turbiwell A 96 250 511 091120 ...
Страница 35: ...A 96 250 511 091120 AMI Turbiwell Installation 35 Degasser G B C Degasser sample inlet Drain Waste B G C ...
Страница 108: ...A 96 250 511 091120 AMI Turbiwell Notes 108 12 Notes ...