A-96.250.511 / 091120
AMI Turbiwell
Installation
30
7
Unscrew the elbow fitting [E] from the inlet of the Flowcontroller.
8
Remove the teflon tape from the threads of fittings [A] and [E]
and wrap them with new teflon tape.
9
Screw the elbow fitting [E] to the inlet of the constant head and
the elbow hose nozzle [A] to the inlet of the flow controller.
10
Install the locking plate [I].
11
Tighten the screw [J].
C
E
Flowcontroller
Elbow fitting
E
C