A-96.250.511 / 091120
AMI Turbiwell
Maintenance
72
6.5.2
Wet Verification
Wet verification can be used instead of verification with a Verikit. It is
performed with a standard with a known turbidity filled into the mea
-
suring chamber instead of the sample.
Note:
When performing a wet verification do not use a standard
below 1 FNU/NTU.
1
Navigate to menu <Maintenance>/
<Verification> and press [Enter].
2
Choose <Other method> and press
[Enter].
3
Stop sample flow.
4
Open the measuring chamber.
5
Open the drain valve if the instru
-
ment is not equipped with an auto
-
matic drain valve.
6
Wait until the measuring chamber is
empty.
7
Seal the overflow [E] with the rubber
stopper [F].
.
8
Close the drainage valve.
9
Fill the measuring chamber with the
standard.
10
Close the measuring chamber.
11
Press [Enter].
3.1
Matching
Maintenance
Set Time 01.01.05 16:30:00
Calibration
Simulation
Verification
3.1.2
Verification
Swan Verikit
Other method
3.1.2.5
- Close inlet valve.
- Empty meas. chamber
Other method
- Open meas. chamber
- Open drainage valve
<Enter> to continue
3.1.2.5
- Plug drain 1 with a
Other method
rubber stopper.
<Enter> to continue
3.1.2.5
- Close Drainage Valve
Other method
- Fill with standard
- Close meas. chamber
<Enter> to continue
Содержание AMI Turbiwell
Страница 1: ...Operator s Manual Firmware V6 23 and higher AMI Turbiwell A 96 250 511 091120 ...
Страница 35: ...A 96 250 511 091120 AMI Turbiwell Installation 35 Degasser G B C Degasser sample inlet Drain Waste B G C ...
Страница 108: ...A 96 250 511 091120 AMI Turbiwell Notes 108 12 Notes ...
Страница 109: ...A 96 250 511 091120 AMI Turbiwell Notes 109 ...