A-96.250.511 / 091120
AMI Turbiwell
Maintenance
62
Possible error messages, see
.
You are prompted to choose between:
[Keep existing]
and
[Save new].
11
Open the measuring chamber.
12
Open the drain valve.
13
Remove the rubber stopper.
14
Rinse the measuring chamber with
sample.
15
Close the measuring chamber.
16
Close the drain valve.
17
Press [Enter].
You are asked whether you want to
match the Verikit. Without a matched
Verikit it is not possible to perform a ver
-
ification. It is also possible to match a
.
18
Start sample flow.
3.32.4
Factor existing
1.00
Save new
Calibration
Factor new
0.92
Keep existing
3.32.5
- Open meas. chamber
- Rinse meas. chamber
Calibration
- Drainage valve open?
- Remove rubber stopper
<Enter> to continue
3.32.5
- Close meas. chamber
Calibration
- Close Drainage Valve
- Drainage valve closed?
<Enter> to continue
3.32.4
- Match Verikit?
Calibration
No
Yes
3.32.5
- Start sample flow
Calibration
<Enter> to finish
Содержание AMI Turbiwell
Страница 1: ...Operator s Manual Firmware V6 23 and higher AMI Turbiwell A 96 250 511 091120 ...
Страница 35: ...A 96 250 511 091120 AMI Turbiwell Installation 35 Degasser G B C Degasser sample inlet Drain Waste B G C ...
Страница 108: ...A 96 250 511 091120 AMI Turbiwell Notes 108 12 Notes ...
Страница 109: ...A 96 250 511 091120 AMI Turbiwell Notes 109 ...