A-96.250.511 / 091120
AMI Turbiwell
Troubleshooting
75
7.
Troubleshooting
7.1. Calibration Errors
Error message:
Deviation too big!! Please check manual.
7.2. Matching Errors
Error message:
Deviation too big!! Please check manual.
7.3. Verification Errors
Error message:
Deviation too big!! Please check manual.
Possible cause
Corrective Action
Wrong formazine stan
-
dard.
Check formazine standard. If
necessary prepare a new formazine
Measuring chamber dirty.
Clean measuring chamber, see
Cleaning the Measuring Chamber,
Possible cause
Corrective Action
Calibration
Perform a new calibration, see
Verification kit dirty.
Clean the verification kit, see
Possible cause
Corrective Action
Verification kit dirty.
Clean the verification kit, see
Verification kit reference
value.
Perform a new matching, see
.
Check if the correct Verikit is used.
Wrong Verikit in use.
Check if the correct Verikit is used.