A-96.250.511 / 091120
AMI Turbiwell
Maintenance
60
Prepare the
20 NTU
formazine
standard
1
Put the pipette into the fast-release pipette pump.
2
Make sure that the pump piston is pushed in completely.
3
Turn the pump wheel until the formazine standard 4000 NTU
reaches the 5 ml level of the pipette.
4
Put the pipette into the volumetric flask and press the fast release
lever until the pipette is empty.
5
Fill the volumetric flask with 1 liter of deionized dilution water.
WARNING
Health hazard
Formazine is harmful to the environment.
At no means recirculate it into the water system.
A
B
C
D
Pipette 5 ml
Fast-release pipette
pump
Rubber stopper
Volumetric flask
A
B
C
D