2.
Product Description
Application
The AMI Turbiwell is used to measure turbidity in potable water, sur
-
face water, effluents and water steam cycles.
The turbidimeter is also suitable for the measurement of other liquids
of which the turbidity correlates with the concentration of a suspend
-
ed solid or an emulsified liquid, e.g. oil in water. See chapter
Calculation, e.g. “Oil in water”, p. 48
for details.
Available
Models
The instrument is available in three different models:
AMI Turbiwell 7027: with an IR LED according to ISO 7027
AMI Turbiwell W/LED: with white light LED (approved, alterna
-
tive method to US EPA 180.1)
AMI Turbiwell Power: variant on steel panel with IR LED
according to ISO 7027
Configurations
and Options
The AMI Turbiwell 7027 and the AMI Turbiwell W/LED are available
in the following configurations:
Automatic or manual drain valve
Sample degasser (option)
deltaT flow sensor (option)
Flowcontroller (option)
On small panel with remote transmitter
The AMI Turbiwell Power is equipped with a flow meter as standard
and available in the following configurations:
Automatic or manual drain valve
Signal
Outputs
Two signal outputs programmable for measured values (freely
scalable, linear or bilinear) or as continuous control output (control
parameters programmable).
Current loop:
0/4–20 mA
Maximal burden:
510 Ω
Third signal output available as an option. The third signal output can
be operated as a current source or as a current sink (selectable via
switch).
Relay
Two potential-free contacts programmable as limit switches for mea
-
sured values, controllers or timer for system cleaning with automatic
hold function. Both contacts can be used as normally open or nor
-
mally closed.
Maximum load: 1 A / 250 VAC
Содержание AMI Turbiwell
Страница 1: ...Operator s Manual Firmware V6 23 and higher AMI Turbiwell A 96 250 511 091120 ...
Страница 35: ...A 96 250 511 091120 AMI Turbiwell Installation 35 Degasser G B C Degasser sample inlet Drain Waste B G C ...
Страница 108: ...A 96 250 511 091120 AMI Turbiwell Notes 108 12 Notes ...
Страница 109: ...A 96 250 511 091120 AMI Turbiwell Notes 109 ...