A-96.250.511 / 091120
AMI Turbiwell
Maintenance
71
Cleaning the
Verification Kit
CAUTION
Damage to optical surfaces
Never use organic agents i. e. alcohol to clean the optical surfac
-
es of the verification kit.
Use a dry cleaning cloth for lenses to clean the optical surfaces
[A]. If necessary, moisten with demineralized water.
If the verification kit is wet, dry it with warm air of max. 50 °C.
Solid
verification kit
Solid verification kit: If the verification is still out of range after clean
-
ing, send the verification kit back to the manufacturer for cleaning
and recertification.
Liquid
verification kit
A
B
Acrylic glass prisms
Housing
A
B
A
B
Acrylic glass windows
Housing
A
B