Chapter 2 Unpacking and Connection
LiQuilaz II E and S Operations Manual
Page 2-11
Communication Options
The LiQuilaz II offers multiple communication options.
Facility Net
Connect using Ethernet or RS-485. When communicating via Ethernet, Facility Net V4.0
or newer is recommended. V3.3 can be used by setting Retro mode in the
LiQuilaz II. This causes the instrument to appear as a HSLIS sensor so that it can be
recognized by the older Facility Net version. Please see
Ethernet communication settings and options must be set via RS-232 setup commands
in
SamplerSight or SamplerSight Pharma
Connect using RS-485. An opto-isolated RS-485 to USB adaptor with a 10ft RS-485
cable to the LiQuilaz II is available (P/N 1000021389).
SamplerSight software can only operate one sensor at a time. If more than one sensor
must be connected via RS-485 via daisy-chain, use FacilityNet software instead.
Modbus TCP
Connect using Ethernet. Communication settings and options must be set via RS-232
setup commands in
, Configuring for Ethernet. Modbus protocol is an
industry standard protocol that is provided for customers who need to interface to a
custom computer program or SCADA system. By default the LiQuilaz II is configured
with Modbus disabled. Use serial command
“set mode 1”
Serial Setup Command List to enable Modbus. The Modbus protocol for the LiQuilaz II
is described in
RS-232
This is primarily a configuration interface to program the LiQuilaz II for the
communication methods mentioned above. An RS-232 adaptor pigtail (P/N PMS-
CD1995) and a USB to RS-232 adaptor (P/N 1000023262) are supplied as accessories to
the LiQuilaz II to facilitate setup. Please see
for serial commands and
, Serial Communications Overview for basic serial
communications information if necessary.