Appendix B RS-485 Communication Protocol
LiQuilaz II E and S Operations Manual
Page B-5
Slow Protocol Commands
Command
Response
Purpose
NOTE:
Commands are case sensitive
CMODE n
RMODE
Select time-based or sample-based sampling mode
CSI n
RSI
Set the sample interval, in seconds, for the time-based
mode
CDT ...
RDT
Set date and time
CTD
RTD ...
Transmit a data report
CPQ
RPQ
Remove the oldest record from queue
CFQ
RFQ
Delete all untransmitted reports
CQC
RQC n s
Return the number of reports on the transmission queue
and the sampling status
CSR
RSR
Software reset
CSS
RSS
Start time-based sampling mode
CTS
RTS
Terminate current sample and stop sampling
CVER
RVER ...
Request LiQuilaz II software version
CSATH ...
RSATH n
Set absolute PHA thresholds
CRATH
RRATH ...
Read absolute PHA thresholds
CSPHA ...
RSPHA n
Set the PHA parameters
CRPHA
RRPHA ...
Read PHA parameters
CSIZE ...
RSIZE n
Set channel sizes
CRSIZE
RRSIZE ...
Read channel sizes
CSTAB ...
RSTAB n
Set table values
CRTAB
RRTAB
Read table values
CADDRESS n
RADDRESS n
Set instrument address