Page G-4
LiQuilaz II E and S Operations Manual
Appendix G Serial Setup Command List
“sta(tus)”
Provide a current list of settings in the unit.
The unit will respond to a “status” command with a serial number
and/or calibration date appended to the bottom if either or both of
those parameters are entered.
See example response below:
LiQuilazII
-
E15P Version:1.08 51
Built: April 27, 2017
MAC Address: 00:60:A6:FF:02:AC
**********Current IP Parameters*************
IP Address: 010.255.002.172
Multicast Address: 224.100.100.001
Net Mask: 255.000.000.000
Gateway: 010.255.000.060
NTP Address: 000.000.000.000
**********After Write IP Parameters*********
IP Address: 010.255.002.172
Multicast Address: 224.100.100.001
Net Mask: 255.000.000.000
Gateway: 010.255.000.060
NTP Address: 000.000.000.000
********************************************
Sample Interval: 5
Connected to: Not Connected
When not connected, the led flashes green
Data Queue set at: 480
Ethernet Connection via PMS Protocol
Serial Num: 12345
Calibrated: 01/31/2017
Mode: Time
Start: On
State: Sampling
Repeat: 0
Tare: 0
“write”
Writes all current non-volatile storage settings into the EPROM.
Table
G-1
Serial setup command list (Continued)