30001:
Modbus Map Version
30002:
Sensor Firmware Version
30003 - 30010:
Product Name
30011:
Flow Rate Scale Factor
30012:
Volume Scale Factor
30013 - 30014:
Flow Rate
30015:
Number of Particle Channels
30016:
Number of Analog Channels
30017 - 30048:
Dual-Register
Representation of the
Particle Channel Sizes in
Nanometers
Appendix C Modbus TCP Protocol
LiQuilaz II E and S Operations Manual
Page C-3
This is a fixed value representing the version number of the
LiQuilaz II Modbus map multiplied by 100. The version can
change if/when the basic structure of the map and/or
processing associated with the structure of the map is
adjusted.
Note: Additions and/or definitions to the spare entries will not
generally require a version change - they will simply be
annotated as to the applicability to that particular instrument.
The same reasoning is applied to unique registers and/or coils
for a particular instrument - they will simply be annotated as
being specific to that device.
This is a value representing the version number of the firmware
multiplied by 100.
Product name entered using ASCII bytes and positioned as
shown in the map.
The power of 10 multiplied to the flow rate and then entered
into registers 30013/30014 and 30222/30223. Used when the
floating-point representation is disabled. The LiQuilaz II uses a
value of 3 to represent a fixed scaling factor of 1000 for flow
rate.
The power of 10 multiplied to the volume and then entered
into registers 30224/30225. Used when the floating-point
representation is disabled. The LiQuilaz II uses a value of 2 to
represent a fixed scaling factor of 100 for volume.
Dual-register representation of the nominal flow rate in
milliliters per minute. Can be integer or float depending on
mode selected. The integer representation uses scale factor
from register 30011.
Processed as a variable but fixed to a value of 16 for LiQuilaz II.
Although the LiQuilaz II will show a fixed value of 16 indicating
that the maximum allowed number of channels is 16, fewer
channels may be configured and the currently unused size
channels will be loaded with a value of zero.
Processed as a variable but fixed to a value of 1 for LiQuilaz II.
The LiQuilaz II allows for a variable number of channels and
variable sizes for each channel. All 16 channels can be read,
and unused size channels will show a value of zero.