Chapter 2 Unpacking and Connection
LiQuilaz II E and S Operations Manual
Page 2-9
Online Version Installation
The LiQuilaz II E20 flow rate is 70 mL/minute ± 10
% and is typically installed directly into
a process line.
The LiQuilaz II E15P flow rate is 20 mL/minute ± 10
%; for online applications 1/4-inch
Flaretek fittings may be specified.
A data cable connects the liquid particle counter unit to a PC running Facility Net,
Pharmaceutical Net, SamplerSight, or SamplerSight-Pharma software that provides
the user interface.
Installation Requirements
Required Items That Are Not Provided
• Liquid Installation Kit
• Tubing, 1/4-inch, Teflon
• Tube-flaring tools
• Communications cable: Ethernet OR
Ethernet Crossover cable (included)
>> To connect the LiQuilaz II (Online version):
1.
Place the particle counter unit where it will have sturdy support and be free of drips
and spray.
2.
Position the PC at the location from which the LiQuilaz II Online will be controlled
and monitored.
3.
Connect a standard IEC AC power cord to the rear panel power module.
4.
Connect a communications cable between the sensor and the data display system.
There are multiple options for communications. See the
2-11.
5.
Cut length of tubing required to connect from the process line to the liquid particle
counter. The Liquid Installation Kit includes 25 ft of 1/4-inch PFA tubing.
6.
Flare the tube end that will connect to the LiQuilaz II.
Figure 2-13
Flared tubing
7.
Connect the flared input tube end to the right Flaretek fitting.