![Spectris Particle Measuring Systems LiQuilaz II E Скачать руководство пользователя страница 82](http://html.mh-extra.com/html/spectris/particle-measuring-systems-liquilaz-ii-e/particle-measuring-systems-liquilaz-ii-e_operation-manual_1344043082.webp)
Appendix D Ethernet Communications Overview
LiQuilaz II E and S Operations Manual
Page D-9
Figure D-5 Internet Protocol (TCP/IP) Properties
window
d.
Record the existing settings on the
Internet Protocol (TCP/IP) Properties
window.
These settings will need to be restored after the instrument communication
session is complete.
e.
Change the settings on the
Internet Protocol (TCP/IP) Properties
window to
use the network address and mask determined above. See
-9 for an example. The
Default gateway
and
DNS server
values should be left
blank.
f.
Click the
OK
button to close the
Internet Protocol (TCP/IP) Properties
window.
g.
Click the
OK
button to close the
Local Area Connection Properties
window.
4.
The PC is now configured.
5.
Attach a network cable between the PC and the instrument.
6.
Use the procedure
Troubleshooting Ethernet Connectivity
D-4 to check
the connection is functioning.
Содержание Particle Measuring Systems LiQuilaz II E
Страница 9: ...viii LiQuilaz II E and S Operations Manual This page is intentionally left blank ...
Страница 15: ...Page 1 4 LiQuilaz II E and S Operations Manual Chapter 1 Getting Started This page is intentionally left blank ...
Страница 105: ...Page H 2 LiQuilaz II E and S Operations Manual Appendix H 㦘㹡㒥㦘 䤓䓸德 侯 This page is intentionally left blank ...