Appendix E Configuring for Ethernet
LiQuilaz II E and S Operations Manual
Page E-5
Set gat(eway) aaa.bbb.ccc.ddd
The gateway address is used when communicating across different networks. This
command sets the gateway address in a form of aaa.bbb.ccc.ddd. Each three-digit
series is a value of 0-255 separated by a period (.) character. Use gateway
000.000.000.000 if no gateway device is available on the network. An example of a valid
gateway address is:
010.255.000.060
.
Set que(ue) x
Enter a data queue size in order to buffer data while the unit is not connected.
The value of
x
can be set anywhere between 1 and 1440. The queue size to enter
depends on the specific application. In general, the value should be made large
enough to minimize data loss when disconnected, but not so large as to create
significant delays in the reception of real-time data when reconnected.
The LiQuilaz II Liquid Particle Counter will:
•
Continue sampling if it was initially started, and then disconnected
•
Queue up to the number of samples defined by the queue size
•
Once full, older data will be overwritten by newer data
•
Unload the queued data in chronological order when reconnected
•
Periodically delete elements of the queue for which Facility Net has received the
data
Set mode x
The unit will run under one of two Ethernet operational modes: PMS Operational Mode
or Modbus Mode. When shipped from the factory, the unit is in the PMS Operational
Mode.
•
When the value of
x
is set to zero (
0
), the unit will reboot into the PMS Operational
Mode.
•
When the value of
x
is set to two (
1
), the unit will reboot into the Modbus Mode.
Select the PMS Operational Mode for use with Facility Net.
Select the Modbus mode to communicate via Modbus TCP (see
Write
Saves the most recent changes.
You must send this command
after setting up the
instrument, or your parameters will be lost when the sensor is de-energized.