Appendix D Ethernet Communications Overview
LiQuilaz II E and S Operations Manual
Page D-5
a.
Is the cable bad?
a.1
Try a different cable.
a.2
If a direct laptop to instrument connection is being done, a cross-over cable
may be required. Cross-over cables are specially marked as such. Particle
Measuring Systems P/N 1000011477 is a cross-over cable.
b.
Is the other end of the cable alive?
Does a known good device work when connected instead?
2.
Verify the device can talk using
PING
.
a.
Open a command shell on a PC, and type
PING aaa.bbb.ccc.ddd
where aaa.bbb.ccc.ddd is the address of the device.
If there is no response, then there is an addressing problem.
Figure D-2
Example for
PING
test
b.
Make sure the PC and the device being talked to are on the same network.
c.
Check the PC’s address by opening a command shell and typing
ipconfig
.
c.1
Is the Mask the same on both the PC and the device?
c.2
Is the network portion of the address the same on both the PC and the
device?