Appendix G Serial Setup Command List
LiQuilaz II E and S Operations Manual
Page G-3
“set mod(e) n”
Set the Ethernet operational mode between PMS or Modbus.
If “n” is 0 then PMS, if “1” then Modbus. The unit will automatically
reboot after the command.
“set mul(ticast)
aaa.bbb.ccc.ddd”
Range Limits
0 – 255
Set Ethernet multicast address.
The values aaa.bbb.ccc.ddd represent decimal entries for each
address octet. Do not use leading zeros. The multicast address is
normally only used by Facility Net. This address must match the
multicast address entered in Facility Net in order to have the full
configuration utilities presented in the configuration dialog. This
setting will be saved after the “write” command is issued.
“set ntp
aaa.bbb.ccc.ddd”
Set Ethernet NTP (Network Time Protocol) address.
The values aaa.bbb.ccc.ddd represent decimal entries for each
address octet. Do not use leading zeros. This setting will be saved
after the “write” command is issued.
“set que(ue) n”
Set the data queue size for use in the PMS and Modbus Operational
modes.
The value for “n” should be between 1 and 1440. The unit will
automatically reboot after the command.
“set sam(ple) n”
Set sample interval into the unit.
The sample interval “n” needs to be a positive integer greater than
or equal to 1 and less than or equal to 28800. The units are in
seconds. The unit will stop sampling if this parameter is changed
during a run state. This setting will be saved after the “write”
command is issued.
“set serdump n”
Set the unit to dump serial data.
If “n” is 1 then serial dump is enabled, otherwise it is disabled. This
setting will be saved after the “write” command is issued. Unlike
the associated “data” command, this feature allows a method by
which the unit can automatically dump data after boot. This
feature is indicated in the status message if enabled.
“set telnet n”
Set the Telnet setting.
If “n” is 1 then the Telnet server is enabled, otherwise disabled.
This can be used for security purposes, the Telnet server is enabled
by default. This setting will be saved after the “write” command is
issued.
“set inst(rument) float n”
Set the Modbus float register representation setting.
If “n” is 1 then the float representation is in effect, otherwise the
integer representation is in effect. This setting will be saved after
the “write” command is issued.
“set inst(rument) ntp n”
Set the NTP setting.
If “n” is 1 then the NTP client is enabled, otherwise disabled. This
setting will be saved after the “write” command is issued.
“set inst(rument) retro n”
Set the retro mode.
If “n” is 1 then the retro mode is enabled, otherwise disabled. This
setting will present the Family Name of “HSLIS” for use by versions
of Facility Net prior to 4.0. This setting will be saved after the
“write” command is issued.
Table
G-1
Serial setup command list (Continued)
Command