Code
Description
Error remedy method
Refer to section
95-C60
A system error was detected
in Disc Drive LSI 1b (Base
Drive). (On side 0)
1.
Acquire the drive/driver log.
2.
Turn off and on the power.
3.
Eject a cartridge.
*1
If the cartridge is not ejected, take the fol-
lowing measures.
• Turn off and on the drive. Press the
EJECT button after startup to eject the
cartridge.
• If the cartridge is not ejected even after
this operation, eject it manually.
4.
Insert the same cartridge.
• If the cartridge cannot be inserted, turn
off and on the drive and then insert the
cartridge again.
• If this error occurs again, take the fol-
lowing actions.
5.
When parts have been replaced, check their
installation status.
6.
Upgrade the drive firmware to the latest ver-
sion.
• If the same error occurs repeatedly,
send the acquired drive/driver log to
your local Sony Sales Office/Service
Center.
“1-8. Location of Main Parts”
“1-10. How to Eject the Car-
tridge (ODS-D280F)”
“1-11. How to Eject the Car-
tridge Manually”
“7. Firmware Upgrade”
95-C61
A servo firmware communi-
cation error was detected in
Disc Drive LSI 1b (Base
Drive). (On side 0)
1.
Eject a cartridge.
2.
Acquire the drive/driver log.
3.
Turn off the drive.
4.
Turn on the drive and insert the same car-
tridge.
• When no error occurs, use the unit as
it is
• When an error occurs, the drive may
be defective.
―
96-907
Head data cannot be read
from the EEPROM in optical
block assembly Ch-2A. (On
side 1)
1.
After the sub base assembly is replaced, take
necessary actions.
2.
Check the sub base assembly installation sta-
tus.
“1-16. Service Action after Re-
placing Blocks and Mounted
Boards”
“4-29. Sub Base Assembly”
“5. Adjustment of Optical Drive
System”
96-90A
A mismatch of OP ID was
detected in the hours meter
data on the Ch-2A side and
the EEPROM data of the op-
tical block assembly Ch-2A.
(On side 1)
1.
After the sub base assembly is replaced, take
necessary actions.
2.
Check the sub base assembly installation sta-
tus.
“1-16. Service Action after Re-
placing Blocks and Mounted
Boards”
“4-29. Sub Base Assembly”
“5. Adjustment of Optical Drive
System”
96-930
There is no response from the
optical block assembly
Ch-2A. (On side 1)
1.
After the sub base assembly is replaced, take
necessary actions.
2.
Check the sub base assembly installation sta-
tus.
“1-16. Service Action after Re-
placing Blocks and Mounted
Boards”
“4-29. Sub Base Assembly”
“5. Adjustment of Optical Drive
System”
96-942
Disc Drive LSI 2a (Base
Drive) cannot write data to
the FLASH (Base Drive).
(On side 1)
1.
After the sub base assembly is replaced, take
necessary actions.
2.
Check the sub base assembly installation sta-
tus.
“1-16. Service Action after Re-
placing Blocks and Mounted
Boards”
“4-29. Sub Base Assembly”
“5. Adjustment of Optical Drive
System”
Continued
*1:
For the ODS-D280F, perform “Eject” operation from the PC (or host device) or press the EJECT button on the rear of the drive to eject
the cartridge.
Note
EJECT button operation or power off/on operation may generate another alarm code.
ODS-D280U/ODS-D280F
2-82
Содержание ODS-D280F
Страница 4: ......
Страница 9: ...Frame Wiring 10 3 Revision History ODS D280U ODS D280F 5 ...
Страница 10: ......
Страница 12: ......
Страница 42: ...Hold the shaded portions Keep away a screwdriver Objective lens Actuator ODS D280U ODS D280F 1 30 ...
Страница 158: ......
Страница 238: ......
Страница 248: ...Possible cause A cartridge is remaining in the unit ODS D280U ODS D280F 6 4 ...
Страница 274: ......
Страница 278: ......
Страница 300: ......
Страница 304: ......
Страница 306: ......
Страница 307: ......