Code
Description
Error remedy method
Refer to section
91-5B7
Failure in Disc Drive LSI 1a
(Base Drive) firmware up-
grade was detected during
startup. (On side 0)
Turn off and on the drive, and then upgrade the
firmware again.
Note
If the drive is turned off during firmware upgrade,
the firmware upgrade may fail.
• If the same phenomenon occurs again, the
drive may be defective.
91-5C5
Communication between
Disc Drive LSI 1b (Base
Drive) and BRIDGE (IC700
internal: BI-353 (ODS-
D280U)/BF-92 (ODS-
D280F) board) cannot be es-
tablished during startup. (On
side 0)
• When the main base assembly or the DR-683
board or the BI-353 board (ODS-D280U)/
BF-92 board (ODS-D280F) has been re-
placed:
Check the installation status.
• When this error occurs after firmware up-
grade:
Turn off and on the drive, and then upgrade
the firmware again.
“1-8. Location of Main Parts”
“1-14. Cautions on Handling the
Optical Block Assembly”
“1-16. Service Action after Re-
placing Blocks and Mounted
Boards”
“4-19. BI-353/BF-92 Board”
“4-30. Main Base Assembly”
“7. Firmware Upgrade”
91-5C7
Failure in Disc Drive LSI 1b
(Base Drive) firmware up-
grade was detected during
startup. (On side 0)
Turn off and on the drive, and then upgrade the
firmware again.
Note
If the drive is turned off during firmware upgrade,
the firmware upgrade may fail.
• If the same phenomenon occurs again, the
drive may be defective.
91-695
Communication between
SYCPU (IC700 internal:
BI-353 (ODS-D280U)/
BF-92 (ODS-D280F) board)
and 0a (Base Drive) cannot
be established during startup.
Or, failure in 0a (Base Drive)
firmware upgrade was detec-
ted during startup. (On side
0)
• When the main base assembly or the DR-683
board or the BI-353 board (ODS-D280U)/
BF-92 board (ODS-D280F) has been re-
placed:
Check the installation status.
• When this error occurs after firmware up-
grade:
Turn off and on the drive, and then upgrade
the firmware again.
“1-8. Location of Main Parts”
“1-14. Cautions on Handling the
Optical Block Assembly”
“1-16. Service Action after Re-
placing Blocks and Mounted
Boards”
“4-19. BI-353/BF-92 Board”
“4-30. Main Base Assembly”
“7. Firmware Upgrade”
91-6A5
Communication between
SYCPU (IC700 internal:
BI-353 (ODS-D280U)/
BF-92 (ODS-D280F) board)
and 0b (Base Drive) cannot
be established during startup.
Or, failure in 0b (Base Drive)
firmware upgrade was detec-
ted during startup. (On side
0)
• When the main base assembly or the DR-683
board or the BI-353 board (ODS-D280U)/
BF-92 board (ODS-D280F) has been re-
placed:
Check the installation status.
• When this error occurs after firmware up-
grade:
Turn off and on the drive, and then upgrade
the firmware again.
“1-8. Location of Main Parts”
“1-14. Cautions on Handling the
Optical Block Assembly”
“1-16. Service Action after Re-
placing Blocks and Mounted
Boards”
“4-19. BI-353/BF-92 Board”
“4-30. Main Base Assembly”
“7. Firmware Upgrade”
91-6B5
Communication between
SYCPU (IC700 internal:
BI-353 (ODS-D280U)/
BF-92 (ODS-D280F) board)
and 1a (Base Drive) cannot
be established during startup.
Or, failure in 1a (Base Drive)
firmware upgrade was detec-
ted during startup. (On side
0)
• When the main base assembly or the DR-683
board or the BI-353 board (ODS-D280U)/
BF-92 board (ODS-D280F) has been re-
placed:
Check the installation status.
• When this error occurs after firmware up-
grade:
Turn off and on the drive, and then upgrade
the firmware again.
“1-8. Location of Main Parts”
“1-14. Cautions on Handling the
Optical Block Assembly”
“1-16. Service Action after Re-
placing Blocks and Mounted
Boards”
“4-19. BI-353/BF-92 Board”
“4-30. Main Base Assembly”
“7. Firmware Upgrade”
91-6C5
Communication between
SYCPU (IC700 internal:
BI-353 (ODS-D280U)/
BF-92 (ODS-D280F) board)
and 1b (Base Drive) cannot
be established during startup.
Or, failure in 1a (Base Drive)
firmware upgrade was detec-
ted during startup. (On side
0)
• When the main base assembly or the DR-683
board or the BI-353 board (ODS-D280U)/
BF-92 board (ODS-D280F) has been re-
placed:
Check the installation status.
• When this error occurs after firmware up-
grade:
Turn off and on the drive, and then upgrade
the firmware again.
“1-8. Location of Main Parts”
“1-14. Cautions on Handling the
Optical Block Assembly”
“1-16. Service Action after Re-
placing Blocks and Mounted
Boards”
“4-19. BI-353/BF-92 Board”
“4-30. Main Base Assembly”
“7. Firmware Upgrade”
Continued
ODS-D280U/ODS-D280F
2-69
Содержание ODS-D280F
Страница 4: ......
Страница 9: ...Frame Wiring 10 3 Revision History ODS D280U ODS D280F 5 ...
Страница 10: ......
Страница 12: ......
Страница 42: ...Hold the shaded portions Keep away a screwdriver Objective lens Actuator ODS D280U ODS D280F 1 30 ...
Страница 158: ......
Страница 238: ......
Страница 248: ...Possible cause A cartridge is remaining in the unit ODS D280U ODS D280F 6 4 ...
Страница 274: ......
Страница 278: ......
Страница 300: ......
Страница 304: ......
Страница 306: ......
Страница 307: ......