Code
Description
Error remedy method
Refer to section
80-EE5
An abnormal value was de-
tected in the measurement of
eccentricity in the optical
block assembly Ch-2B. (On
side 1)
1.
Eject a cartridge.
2.
Insert again the ejected cartridge.
• When no error occurs, use the unit as
it is.
• When an error occurs, go to step 3) in
Measures.
3.
Insert another cartridge.
• In the case no error occurs:
The ejected cartridge is likely to be
faulty.
• In the case the error occurs:
Drive trouble is likely.
4.
Check the carry loader assembly installation
status.
Note
If the carry loader assembly is abnormally
installed, disc cannot be chucked normally
and the eccentricity may increase.
80-F65
An abnormal value was de-
tected in the measurement of
eccentricity in the optical
block assembly Ch-3A. (On
side 1)
1.
Eject a cartridge.
2.
Insert again the ejected cartridge.
• When no error occurs, use the unit as
it is.
• When an error occurs, go to step 3) in
Measures.
3.
Insert another cartridge.
• In the case no error occurs:
The ejected cartridge is likely to be
faulty.
• In the case the error occurs:
Drive trouble is likely.
4.
Check the carry loader assembly installation
status.
Note
If the carry loader assembly is abnormally
installed, disc cannot be chucked normally
and the eccentricity may increase.
80-FE5
An abnormal value was de-
tected in the measurement of
eccentricity in the optical
block assembly Ch-3B. (On
side 1)
1.
Eject a cartridge.
2.
Insert again the ejected cartridge.
• When no error occurs, use the unit as
it is.
• When an error occurs, go to step 3) in
Measures.
3.
Insert another cartridge.
• In the case no error occurs:
The ejected cartridge is likely to be
faulty.
• In the case the error occurs:
Drive trouble is likely.
4.
Check the carry loader assembly installation
status.
Note
If the carry loader assembly is abnormally
installed, disc cannot be chucked normally
and the eccentricity may increase.
8F-E00
Startup adjustment or OPC
cannot be finished in optical
block assembly Ch-2A. (On
side 1)
1.
Eject a cartridge.
2.
Insert again the ejected cartridge.
• When no error occurs, use the unit as
it is.
• When an error occurs, go to step 3) in
Measures.
3.
Insert another cartridge.
• In the case no error occurs:
The ejected cartridge is likely to be
faulty.
• In the case the error occurs:
Drive trouble is likely.
―
Continued
ODS-D280U/ODS-D280F
2-65
Содержание ODS-D280F
Страница 4: ......
Страница 9: ...Frame Wiring 10 3 Revision History ODS D280U ODS D280F 5 ...
Страница 10: ......
Страница 12: ......
Страница 42: ...Hold the shaded portions Keep away a screwdriver Objective lens Actuator ODS D280U ODS D280F 1 30 ...
Страница 158: ......
Страница 238: ......
Страница 248: ...Possible cause A cartridge is remaining in the unit ODS D280U ODS D280F 6 4 ...
Страница 274: ......
Страница 278: ......
Страница 300: ......
Страница 304: ......
Страница 306: ......
Страница 307: ......