LENS Integrated System     

Operations/Service Manual

Precautions

• 

Hazardous high voltage and energy are present at the output 

and in the internal circuitry of this unit . 

• 

When inspecting the light cable, NEVER aim the light cable at 

the light source . Retinal damage may occur .

• 

Connect only accesory items that have been specified as part 

of the System or have been specified as compatible . Use of the 

accessory, transducer, or cable with equipment and systems 

other than those specified may result in increased emissions or 

decreased immunity of the equipment or system .

• 

During operation, avoid prolonged contact of the scope tip to 

patient tissue or flammable materials . The scope tip may reach 

high temperatures due to high intensity light transmission .

• 

Do not leave the operating light cable on a patient or the 

drapes . Failure to observe this warning may result in burns to 

the patient and/or the surrounding drapes .

• 

Do not use this device in oxygen rich environments (>25%) .

• 

To avoid the risk of electric shock, this equipment must only be 

connected to a supply mains with protective earth .

• 

No modification of this equipment is allowed

• 

Do not spill liquids on the control unit . If liquid enters the unit 

through the vents, permanent damage may result .

If using the Wi-Fi enabled control unit

• 

Use of this equipment adjacent to or stacked with other 

equipment should be avoided because it could result in 

improper operation . If such use is necessary, this equipment 

and the other equipment should be observed to verify that they 

are operating normally .

• 

Use of accessories, transducers, and cables other than those 

specified or provided by the manufacturer of this equipment 

could result in increased electromagnetic emissions or 

decreased electromagnetic immunity of this equipment and 

result in improper operation .

• 

Portable RF communication equipment (including peripherals 

such as antenna cables and external antennas) should be 

used no closer that 30 cm (12 inches) to any part of the LENS 

Integrated System control unit, including cables specified by 

the manufacturer . Otherwise, degradation of the performance 

of this equipment could result .

Precautions

U.S. Federal law restricts this device to sale by or on the order  

of a physician.

•  Prior to each use, inspect the device to ensure it is functioning 

properly and is not damaged. Do not use a damaged device.

•  Do not allow fluids to enter the device. Damage can result.
•  Electrical safety testing should be performed by a biomedical 

engineer or other qualified person.

•  All products used in conjunction with this control unit, such as video 

monitors, video printers, etc., are likely to have an inherent leakage 

current at their video connections. Connection to this control unit 

likely will increase the leakage current at the control unit’s chassis.

•  When looping the video output to more than one piece of 

equipment, the termination switch on the last looping piece of 

equipment should be in the 75U (ohm) or the ON position if the 

equipment provides that option.

•  Do not plug the camera cable connector into the control unit if 

wet. Moisture on the card edge of the camera cable will damage 

the circuitry and void the warranty. Ensure that the card edge is 

completely dry before plugging the camera head into the camera 

control unit.

•  This equipment has been tested and found to comply with the limits 

for a Class A digital device, pursuant to Part 15 of the FCC Rules. 

These limits are designed to provide reasonable protection against 

harmful interference in a industrial installation. This equipment 

generates uses and can radiate radio frequency energy and, if 

not installed and used in accordance with the instructions, may 

cause harmful interference to radio communications. However, 

there is no guarantee that interference will not occur in a particular 

installation. If this equipment does cause harmful interference to 

radio or television reception, which can be determined by turning 

the equipment off and on, the user is encouraged to try to correct 

the interference by one of the following measures:

 

– Reorient or relocate the receiving antenna.

 

– Increase the separation between the equipment and receiver.

 

– Connect the equipment into an outlet on a circuit different from 

that to which the receiver is connected.

 

– Consult the dealer or an experienced radio/TV technician for 

help.

•  This device complies with Part 15 of the FCC Rules. Operation is 

subject to the following two conditions: (1) This device may not 

cause harmful interference, and (2) this device must accept any 

interference received, including interference that may cause 

undesired operation.

•  Maintain a minimum separation distanceof 20 cm between the 

antenna and the patient’s body.

•  Any changes or modifications not expressly approved by the party 

responsible for compliance could void the user’s authority to operate 

this equipment.

•  Environmental Protection: This equipment contains electronic printed 

circuit assemblies. At the end of the useful life of this equipment, it 

should be disposed of in accordance with any applicable national or 

institutional policy relating to obsolete electronic equipment.

•  Products may wear over time as a result of normal use. Periodic 

retesting of the leakage current is recommended.

•  Do not operate at line voltages other than those stipulated on the 

back of the unit.

•  Handle the unit with care. If the unit is dropped or damaged in any 

way, it must be returned immediately for service.

•  To avoid loss of or damage to the control unit, remove the antenna 

from the Wi-Fi enabled control unit prior to transport.

•  Disconnect the power cord before cleaning the unit.
•  Ensure that the unit is installed in an environment in which the vents 

and fans are not obstructed. Insufficient ventilation can cause the 

unit to overheat and shut down.

•  In the event that the unit overheats and shuts down but the cooling 

fans are still running, do not turn off or unplug the unit. Continue to 

run the fans to prevent further damage to the unit.

•  All products used in conjunction with this CCU, such as video 

monitors, video printers, etc., are likely to have an inherent leakage 

current at their video connections. Connection to this CCU likely will 

increase the leakage current at the CCU’s chassis.

Содержание LENS Integrated System

Страница 1: ...LENS Integrated System Operations Service Manual...

Страница 2: ......

Страница 3: ...general waste Caution See Instructions for Use Follow instructions for use CE Mark Select button Danger Non ionizing electromagnetic radiation Brightness Power plug e Environmentally friendly use per...

Страница 4: ...uide 20 Turn the Camera Control Unit On 20 Connect the iPad 21 Inspect the System Components 21 Electrical Connections 21 Inspect the Fiber Optic Light Cable 21 Safety and Service Life Checklist 22 Op...

Страница 5: ...ction with the LENS Integrated System Wi Fi version is indicated for capture of still and motion pictures patient file management and limited redundant control of the LENS Control Unit within articula...

Страница 6: ...ion Do not plug the camera cable connector into the control unit if wet Moisture on the card edge of the camera cable will damage the circuitry and void the warranty Ensure that the card edge is compl...

Страница 7: ...parately and the optional LENS App It is configured to be connected as a stand alone system on an endoscopic cart or tower The System can be used in conjunction with compatible high definition flat pa...

Страница 8: ...is designed to accept the Smith Nephew Wolf Light Guide end fittings 9 Camera Cable Connector receptacle Compatible Smith Nephew camera heads can be connected via the card edge connector Be sure the...

Страница 9: ...ow control of peripheral devices such as printers and image capture devices from the camera head 5 Fuse Holder The clip holds a dual fuse See the Technical Specifications section of this manual 6 AC M...

Страница 10: ...a customized procedure The two functions are activated by a short press 1 0 seconds and a long press 1 0 seconds These buttons can be customized based on the user s needs Refer to the Button Settings...

Страница 11: ...utton settings refer to the Button Settings section of this manual Function Effect BRIGHTNESS Control the luminance level of the video output ENHANCEMENT Enhance the sharpness of the displayed video Z...

Страница 12: ...ce are connected to the System a patient name b number of images and videos recorded c current procedure name Note When both an iPad and a USB are connected the iPad information will appear on the OSD...

Страница 13: ...be placed on a platform or cart shelf equipped with a screw mount follow these instructions to secure the control unit to the platform or shelf 1 Center the control unit on the shelf and ensure that...

Страница 14: ...and HD SDI connectors BNC SDI composite To connect push in and turn BNC HDSI To connect push in and turn Note BNC SDI and BNC HDSI connectors appear identical For the highest quality image be sure to...

Страница 15: ...ol unit s rear panel 3 To record still or video images from the camera connect the App to the LENS Wi Fi network or a insert a USB storage device in one of the USB ports on the control unit Refer to t...

Страница 16: ...16 10601034 Rev B LENS Integrated System Operations Service Manual Installation Figure 5 Recommended setup configuration Wi Fi enabled System...

Страница 17: ...nagement system is compatible only with 1080i output Refer to the Advanced Settings section of this manual for information about how to configure input and output settings 4 To record still images fro...

Страница 18: ...Sistema integrado Integreret system Integrert system Endoscopy Smith Nephew Inc 150 Minuteman Road Andover MA 01810 USA T 1 978 749 1000 F 1 978 749 1108 Customer Service 1 800 343 5717 HDSDI 2 HDSDI...

Страница 19: ...rimary left and secondary right inputs Color Explanation Orange Ready White Input is not ready Red Recording or busy For example in Figure 7 the left side of the icon is orange which indicates that th...

Страница 20: ...technique in accordance with standard operating room procedures 1 Attach a videoscope or coupler with a sterilized direct view endoscope to the camera head 2 Attach the fiber optic light guide from th...

Страница 21: ...inspected for damage which can negatively impact the system s performance Inspection should include all equipment to be used in surgery including cables and peripheral devices CAUTION Prior to each u...

Страница 22: ...the Wi Fi is connected Check front panel to ensure that all buttons and connectors are functioning properly Check the OSD status bar to confirm that the light is turned on and detected Check the OSD s...

Страница 23: ...Use the Up and Down arrows on the control unit or the left and right camera head buttons to navigate to the desired feature or function Note Settings changed directly from the MAIN MENU are temporary...

Страница 24: ...reen Figure 18 Refer to the Customize a Procedure section of this manual for information about how to customize or create a procedure Figure 18 To exit the SELECT PROCEDURE menu highlight the Menu Acc...

Страница 25: ...mage storage devices For information regarding how to capture still images from the App please refer to LENS Application Instructions for Use REF 10601295 For information about how to customize button...

Страница 26: ...eset Factory Defaults icon on the SYSTEM CONFIGURATION MENU screen Note that if an administrator password has been set the System will ask the user for the administrator password before allowing passw...

Страница 27: ...ROCEDURES SUB MENUS PROCEDURE SELECTION CUSTOMIZE PROCEDURE SET DEFAULT PROCEDURE CAMERA BUTTON MAPPINGS N STANDBY POWER SELECT PROCEDURE Y EXIT MENU ACCESS FRONT PANEL CAMERA HEAD WHITE BALANCE SYSTE...

Страница 28: ...urn off the power switch on the tower To print From the LENS App refer to the App s Instructions for Use REF 10601250 for information From the 660HD Image Management System refer to that product s man...

Страница 29: ...icon on the SYSTEM CONFIGURATION MENU screen to open the CHANGE PASSWORD screen Figure 22 PASSWORD REQUIRED is blinking The default is set to NO Figure 22 2 Select PASSWORD REQUIRED The current setti...

Страница 30: ...hout saving any changes navigate to CANCEL and press Select to return to the SYSTEM CONFIGURATION MENU screen Note Do not lose the administrative password This will prevent the recovery of passwords t...

Страница 31: ...and return to the SYSTEM CONFIGURATION MENU screen Select NO to retain previously saved settings and exit the RESET FACTORY DEFAULTS screen without making any changes The System returns to the SYSTEM...

Страница 32: ...channels available in the System channels 0 16 If the System performance seems to be slowing or otherwise degrading change to a different network channel to try to resolve the problem For more inform...

Страница 33: ...gh the available options Figure 35 Press Select to activate the desired option POWER LOSS ALWAYS NEVER Figure 35 To configure the camera head plugin The HEAD PLUGIN option allows the user to customize...

Страница 34: ...ess BRIGHTNESS ENHANCEMENT ZOOM TOP PORT BOTTOM PORT PAUSE VIDEO ALT PAUSE ALT VIDEO ALT FRAME FRAME CAPTURE LIGHT MODE STRETCH VIDEO CAPTURE LAST IMAGE MENU ACCESS WHITE BALANCE Figure 38 Refer to Ta...

Страница 35: ...TNESS blinks press Select to select it The current setting will blink 2 Use the Up and Down arrows on the control unit or the left and right camera head buttons to scroll through the BRIGHTNESS level...

Страница 36: ...S screen ELC Settings ELC Electric Light Control Settings allows the user to customize the response of the ELC The recommended setting is AUTO which is also the default setting It can also be set to s...

Страница 37: ...cessive heat Note It is important to periodically inspect and clean the fan grill on the rear of the unit and the exhaust holes on the side of the unit The most effective method is to use a vacuum wit...

Страница 38: ...ing agent off of the lenses The use of an antifogging agent is not recommended Anti fogging agent had been applied to the lenses Increase the Enhancement setting Use the ENHANCEMENT button on the came...

Страница 39: ...to factory Color phase or hue were incorrectly customized in the procedures Unplug and plug in the camera head and perform white balance If this works and is repeatable call tech support System AC Mai...

Страница 40: ...afety requirements These tests should be conducted in accordance with specifications IEC 60601 1 2005 CORR 1 2006 CORR 2 2007 AM1 2012 IEC 60601 1 2012 Third Edition IEC 60601 2 18 2009 Third Edition...

Страница 41: ...USB devices have been validated for use with the System 4x8 GB USB Sticks Transcend CorsairT ImationT Kingston 4x16 GB USB Sticks ADATAT Mushkin Silcon PowerT generic Team 4x32 GB USB Sticks Corsair P...

Страница 42: ...ystem German 72204846 LENS Integrated System Spanish 72204847 LENS Integrated System French 72204855 LENS Integrated System Italian 72204849 LENS Integrated System Swedish 72204850 LENS Integrated Sys...

Страница 43: ...nd Mobile RF Communications Equipment and the Smith Nephew LENS Integrated System The LENS Integrated System is intended for use in an electromagnetic environment in which radiated RF disturbances are...

Страница 44: ...Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 2 kV line to P E 1 kV line to line 2 kV line to P E 1 kV line to line Mains power quality should...

Страница 45: ......

Страница 46: ......

Страница 47: ......

Страница 48: ...ate of invoice Refer to the current Smith Nephew Product Catalog or contact Smith Nephew Customer Service for specific warranty information Service Replacement Program Smith Nephew offers a 24 hour Se...

Отзывы: