17
10601034 Rev. B
LENS Integrated System
Operations/Service Manual
Installation
Recommended System Configuration, No Wi-Fi
This section explains the recommended equipment connections for
optimum performance. Perform this procedure each time that the
System is installed or reconfigured.
Required Equipment:
• LENS Camera Control Unit (REF 72204685)
• HD Flat Panel Display
Optional Equipment:
• Smith & Nephew 660HD Image Management System: required for
connection to hospital system (PACS/EMR through the Integration
Broker device)
• USB storage device. The USB can be connected to either the USB
port on the front panel or the port on the rear panel of the control
unit.
Procedure
Note: For best image, use a HD-DVI signal (control unit output to
monitor input). A direct connection between the control unit and a high
definition monitor will always produce the highest quality image.
Note: It is strongly suggested that a minimum of two outputs be
connected at all times to provide a back-up signal if necessary.
1. Connect the supplied hospital-grade power cord to the power
receptacle on the control unit’s rear panel.
2. Connect the supplied hospital-grade power cord to the power
receptacle on the 660HD Image Management System and plug it in.
3. Connect the HD-SDI output from the control unit to the HD-SDI
input of the 660HD Image Management System. Then connect the
660HD HD-SDI output to the flat panel display. The 660HD image
management system is compatible only with 1080i output. Refer to
the “Advanced Settings” section of this manual for information about
how to configure input and output settings.
4. To record still images from the camera, connect the 3.5 mm mini
phono plug from one of the control unit’s peripheral controls
outputs (Peripheral 1 or 2, or top port or bottom port), to the image
management system’s Still Capture IN port.
Note: The camera head buttons must be mapped so that they will
trigger the peripheral ports. Refer to the “Button Settings” section
of this manual for information about how to map the camera head
buttons.
5. To record live video images from the camera, connect the 3.5 mm
mini phono plug from one of the control unit’s peripheral controls
outputs (Peripheral 1 or 2, or top port or bottom port) to the image
management system’s Video Capture IN port.
6. Connect the supplied hospital-grade power cord to the power
receptacle on the flat panel display and plug it in.
7. If the flat panel display has a termination switch, set it to the 75U or
ON position.
8. After connecting all components, including the camera head, turn
the video components on using the appropriate power switches.
Configure the peripheral ports
The default settings for the peripheral ports on the rear of the control
unit are FRAME CAPTURE (Peripheral 1, or the top port) and VIDEO
CAPTURE (Peripheral 2, or the bottom port. Each of the peripheral
control ports can be confitgured for FRAME CAPTURE, GENERAL
PURPOSE, or VIDEO CAPTURE. For information on how to customize the
peripheral ports, refer to the “Configure Peripheral Ports” section of this
manual.