LENS Integrated System
Operations/Service Manual
Cleaning
Cleaning
To Clean the LENS Integrated Control Unit
CAUTION: Disconnect the power cord before cleaning the light source.
• Prior to cleaning the equipment, turn off the control unit and
disconnect the camera head.
• Wipe down the control unit with a clean, dry cloth after every
procedure.
• Use a damp cloth or sponge to remove dirt or debris. Avoid getting
liquid into the side vents, and never immerse the control unit in any
solution.
• Wipe down the control unit with alcohol or a neutral pH cleaner.
• After cleaning, store the equipment in a cool, dry place out of direct
sunlight or excessive heat.
Note: It is important to periodically inspect and clean the fan grill on the
rear of the unit and the exhaust holes on the side of the unit. The most
effective method is to use a vacuum with a soft brush attached. A soft,
damp cloth may also be used to remove any accumulations in these
areas.
To Clean and Sterilize the Camera Head
Refer to the Instructions for Use provided with the LENS Camera Head
(REF 10601250).
Note: It is important to select the appropriate sterilization method for
each type of equipment.
For sterilization of a light guide, refer to the “Sterilization” section of that
light guide’s Instructions for Use.
For sterilization of a coupler, refer to the “Sterilization” section of that
coupler’s Instructions for Use.
For sterilization of an endoscope, refer to the “Sterilization” section of
that endoscope’s Instructions for Use.